360142-D/D
25
i.Series® Models
Rotate the spring using a J-hook tool then insert pin (left-hinged door shown).
11
Switch AC ON/OFF switch ON. Switch battery switch ON. Close the lower front control panel.
12
Touch
Mute
to disable the high temperature alarm while freezer reaches operating temperature.
13
Verify the door is level and the hinges operate smoothly and the door seals tightly.
6.9
Stacked Undercounter Units
CAUTION
►
For a stacked configuration, both units must have leveling feet installed.
►
The back brace bars and front stabilizing brackets must be installed [PN 400821-1
(blue) or 400821-2 (stainless steel)].
►
When stacking a refrigerator and freezer (104 and/or 105 models), place the heavier
unit on the bottom.
►
Do not open multiple, loaded drawers or baskets at the same time.
Call Helmer or your distributor for more information on the stacking kit, and for methods to secure both
units to the wall and/or the floor.