Temperature Settings
Temperature setpoint values are programmed at the factory. Setpoints can be viewed and changed
through the i.C³ monitoring and control system. To view temperature setpoints, touch
Home
,
i.C³ APPS
,
Settings
. Details for each setpoint are displayed.
Temperature Controller Programs screen.
NOTE
Default chamber temperature setpoint is -30.0 °C
Change the setpoint if:
►
Your organization requires a chamber temperature other than -30.0 °C.
NOTE
If the Settings screen is password protected or if viewing settings for the first time, enter
factory default password of “1234”.
Perform the following:
1
Touch
i.C³ APPS
,
i.C³ Settings
.
2
Enter the Settings password.
3
Touch
Temperature Setpoints
.
4
Touch
+
or
–
on the
Temperature Setpoint
spin box.
►
The setpoint is the temperature at which the freezer operates.
5
Touch
+
or
–
on the
Hysteresis Setpoint
spin box.
►
The hysteresis setpoint is the allowable fluctuation in temperature, relative to the temperature setpoint.
►
A lower hysteresis setpoint will limit the temperature variation to a smaller range.
►
A higher setpoint will allow the temperature to vary across a larger range.
NOTE
Hysteresis is factory-preset and should not be changed unless directed by Helmer
Technical Service.
6
Touch
+
or
–
on the
Delay on Start-Up
spin box.
►
Compressor startup is delayed to allow the i.C³ monitoring and control system to start first.
7
Touch
+
or
–
on the
Duty Cycle of Control Relay during Probe Failure
spin box.
►
The duty cycle is the percentage of time the compressor will run in the event of a temperature
control probe failure.
Setting
Initial Factory Value
Temperature Setpoint
-30.0 °C
Hysteresis Setpoint
2.0 °C
Delay on Start-Up
2 minutes
Duty Cycle of Control Relay during Probe Failure 100%
Defrost Time
15 minutes