360142-D/D
30
i.Series® Models
7.3.3
Calibrate Compressor and Evaporator Probe
The compressor and evaporator temperature probes have been factory-calibrated. Calibration settings
should not be changed unless directed by Helmer Technical Service.
7.3.4
Factory Default Settings
Settings listed below may be simultaneously returned to factory default values.
NOTE
The factory default settings may not be the same as the settings that were factory-
calibrated before the freezer was shipped.
Setting
Restored Value
Home Screen Application Icons
i.C³ APPS, Temperature Alarm Test, Temperature
Graph, Information Logs, Download
Display Brightness
High (3 symbols)
Password (for Settings screen)
1234
Sounds
On
Alarm Volume
9
Alarm Tone
On
Temperature Calibration Values
Values previously entered during setup
Unit ID
Serial number entered at factory
Date Format
MM/DD/YYYY
Day
Not affected (maintained in real-time clock)
Month
Year
Time Format
12-hour
Minute
Not affected (maintained in real-time clock)
Hour
AM/PM
Language
Language previously selected during setup
Temperature Units
°C
Password Protection (for Settings screen)
On
Temperature Graph Screensaver
On
Access Control (optional) as Home Page
On
High Temperature Alarm Setpoint
-20.0 °C
High Temperature Alarm Time Delay
0 minutes
Low Temperature Alarm Setpoint *
-35.0 °C
Low Temperature Alarm Time Delay
0 minutes
Power Failure Alarm Time Delay
1 minute
Probe Failure Alarm Time Delay
0 minutes
Door Open (Time) Alarm Time Delay
3 minutes
Compressor Temperature Alarm Setpoint
50.0 °C
Compressor Temperature Alarm Time Delay
0 minutes
Chamber Setpoint
-30.0 °C
Chamber Hysteresis
2.0 °C
Delay on Start-Up
2 minutes