360142-D/D
31
i.Series® Models
Setting
Restored Value
Control Relay Probe Failure Duty Cycle
100%
Defrost Event #1 On/Off
Off
Defrost Event #1 Start Time
12:00 AM
Defrost Event #2 On/Off
On
Defrost Event #2 Start Time
8:00 AM
Defrost Event #3 On/Off
Off
Defrost Event #3 Start Time
4:00 PM
Defrost Event #4 On/Off
On
Defrost Event #4 Start Time
6:00 PM
Defrost Time/Defrost Safety Operation Time
15 minutes
7.3.5
Restore Factory Default Settings
1
Touch
Home
,
i.C³ APPS
,
Settings
,
Restore Factory Settings
.
2
A “Are you sure you want to restore factory settings?” message appears. Do one of the following:
►
Touch
Yes
. The message screen closes and factory settings are restored.
►
Touch
No
. The message screen closes and factory settings are not restored.
7.3.6
Change Factory Settings
Several of the freezer operating parameters are configured at the factory. The settings listed below are
set at the factory, and may be changed at the direction of Helmer Technical Service.
Setting
Description
Temperature Controller Page
Enable or disable the temperature controller screen
Factory settings may be viewed and changed. Contact Helmer Technical Service to verify if changing
factory settings is necessary, and for instructions in accessing Factory Settings screen.