360142-D/D
33
i.Series® Models
Cancel the test:
1
Touch
Home
,
i.C³ APPS,
Temperature Alarm Test
.
2
Touch
Cancel Test
.
NOTE
When cancelling an automatic test, the message indicating the test is in progress clears
immediately. If a setpoint was reached before the test was cancelled, the alarm activates
and clears as described earlier.
7.4.2
Manual Chamber Alarm Test
NOTE
Before testing alarms, protect items in freezer from extended exposure to adverse
temperature.
Obtain:
►
glass with 4 oz. (120 mL) of product simulation solution (1:1 ratio of water to propylene glycol or
equivalent low-temperature fluid).
NOTE
Temperature probes are fragile; handle with care.
Test the high alarm:
1
Identify setting for high alarm setpoint.
2
Place the glass of product simulation solution in the freezer.
3
When the product simulation solution has stabilized at the chamber temperature, remove the solution
from the freezer.
4
Remove the monitor probe from the probe bottle and insert into the product simulation solution.
5
Observe the temperature on the i.C³ display at which the high temperature alarm sounds.
6
Compare the temperature at which the alarm sounds to the high alarm setpoint. If values do not
match, refer to
Section II, Item
9
(Troubleshooting).
7
Remove probe from product simulation solution.
8
Place monitor probe in probe bottle, immersing it at least 2” (50 mm).
7.4.3
Power Failure Alarm Test
NOTE
During a power failure, the power failure alarm sounds and the battery provides power to
the monitoring system.
1
Change Power Failure delay setting to 0 minutes.
a
Touch
Home
,
Settings
,
Alarm Settings
.
b
Touch
+
or
–
on the Power Failure spin box to change the value to 0.
2
Switch AC ON/OFF switch OFF. Power failure alarm will activate immediately.
3
Switch AC ON/OFF switch ON. Power failure alarm will clear and audible alarm will cease.
4
Change Power Failure time delay to the original setting.
7.4.4
Door Open Alarm Test
1
Change Door Open (Time) delay setting to 0 minutes.
a
Touch
Home
,
Settings
,
Alarm Settings
.
b
Touch
+
or
–
on the Door Open (Time) spin box to change the value to 0.
2
Open door. Alarm will activate immediately.
3
Close door. Alarm will clear and audible alarm will cease.
4
Change the Door Open (Time) setting to the original setting.