i.Series® Models
8.3
Check Monitoring System Battery
On all i.C³ screens, the Battery icon will appear in the header bar when the system is running on battery
power and the screen brightness will automatically be reduced. The monitoring system will automatically
disable some features to extend battery life.
Check the battery:
1
Switch AC ON/OFF switch OFF.
a
Screen should continue to display information with reduced brightness.
b
Battery icon will appear on the screen.
c
If the display is blank, replace battery.
2
Switch AC ON/OFF switch ON.
NOTE
Use a battery which meets manufacturer’s specifications
8.4
Clean the Freezer
8.4.1
Condenser Grill
!
CAUTION
Disconnect freezer from AC power when cleaning the condenser grill.
In environments where freezer is exposed to excessive lint or dust, condenser grill may require cleaning
more frequently than stated in preventive maintenance schedule.
Clean the condenser grill using a soft brush and a vacuum cleaner.
8.4.2 Exterior
Clean glass surfaces with soft cotton cloth and glass cleaner. Clean exterior surfaces with soft cotton
cloth and non-abrasive liquid cleaner.
8.4.3 Interior
Clean painted surfaces with mild detergent. Clean stainless steel surfaces with a general-purpose
laboratory cleaner suitable for stainless steel.
8.4.4
Door Gaskets
Clean with soft cloth and mild soap and water solution.
8.4.5
Clean and Refill Probe Bottle
Obtain:
►
Fresh water-bleach solution (not provided)
►
1:9 ratio of bleach to water
►
Bleach is 5% solution of commercial sodium hypochlorite (NaOCl)
►
Equivalent oxidizing cleaner/disinfectant approved by your organization may be substituted
►
4 oz (120 mL) of product simulation solution per bottle
►
Solution is a 1:1 ratio of water to propylene glycol (or equivalent low-temperature fluid)
Clean and refill bottle:
1
Remove probe from bottle.
2
Remove bottle from bracket.
3
Clean bottle with water-bleach solution.
4
Fill bottle with 4 oz (120 mL) of product simulation solution.