Horizon Series™ Models
13
Settings
Through the Horizon Series monitoring and control system, current settings may be viewed and changed.
NOTE
►
Control Sensor Offset and Hysteresis settings are factory-preset and should not be
changed unless directed by Helmer Technical Service.
►
Changing temperature settings affects operation of the freezer. Do not change
settings unless instructed in product documentation or by Helmer Technical Service.
13.1
Monitor and Controller Interface
A B D F G H J K L
C E I
Label Description
Function
A
HIGH TEMP lamp
Indicates when the freezer is in a high temperature alarm condition.
Also indicates high alarm temperature setpoint is being changed.
B
LOW TEMP lamp
Indicates when the freezer is in a low temperature alarm condition.
Also indicates low alarm temperature setpoint is being changed.
C
DOOR ALARM lamp
Indicates when the door is open.
D
COOL lamp
Indicates the compressor is running.
E
SET lamp
Indicates when temperature setpoint or alarm setpoint is being changed.
F
Display
Displays real-time temperature information, setpoints, and alarms.
G
SEL button
Toggles between alarm monitor and control modes.
H
CONTROL lamp
Indicates when the reading from the control probe is displayed.
I
MONITOR lamp
Indicates when the display is showing temperature readings from
the monitor probe. Also indicates when alarm setpoints are being
changed.
J
UP ARROW button
Increases a temperature setting.
K
DOWN ARROW /
MUTE ALARM button
Decreases a temperature setting. Also mutes the audible alarm for five
minutes.
L
SET button
Allows settings to be selected, prior to changing settings.
NOTE
The Alarm Disable key switch disables all audible alarms. This switch does not affect
alarm lamps or signals sent through the remote alarm interface.