360142-D/D
74
Horizon Series™ Models
Enter the new offset value:
1
Press and hold the
Up
and
Down Arrows
simultaneously for 3 seconds to enter program mode.
2
The display will show .C or .F to indicate Celsius or Fahrenheit.
3
Press
SEL
until only the MONITOR lamp flashes.
4
Hold
SET
, then press
Up
or
Down Arrow
to change the monitor offset.
5
Release
SET
button. The new setting is saved.
6
Press and hold
Up
and
Down Arrows
simultaneously for 3 seconds to exit program mode.
13.6.2
Control Sensor Offset
►
Factory-preset. Varies for each freezer.
Determine control sensor offset:
NOTE
►
Control Sensor Offset is factory-preset and should not be changed. Contact Helmer
Technical Service for instructions regarding changing the Control Sensor Offset.
►
Steps in Item 13.6.1 must be performed to ensure the monitor display is calibrated and
accurate prior to adjusting the Control Sensor Offset.
1
View and record the Freezer Setpoint. (Reference
Section III
,
Item
13.3
)
2
Allow the unit to run with calibrated monitor temperature for several compressor cycles, and record
the average monitor temperature
.
3
View and record the current Control Offset value.
4
Subtract the Freezer Setpoint from the average monitor temperature and record the difference.
5
Add the current Control Offset value to the recorded difference determined in the previous step to
establish the new Control Offset value.
EXAMPLE
1
Freezer Setpoint is -30.0
2
Average monitor temperature is -31.2
3
Current Control Offset is 0.3
4
Subtract: -31.2 - (-30.0) = -1.2; difference between average temperature and setpoint
5
Add 0.3 + (-1.2) = -0.9; new Control Offset value
Enter the new offset value:
1
Press and hold the
Up
and
Down Arrows
simultaneously for 3 seconds to enter program mode.
2
The Display will show .C or .F to indicate Celsius or Fahrenheit.
3
Press
SEL
until only the CONTROL lamp flashes.
NOTE
►
Ensure Control Sensor Offset is being changed, and not Hysteresis.
►
Control Sensor Offset and Hysteresis have the same visual indicator.
4
Hold
SET
, then press
Up
or
Down Arrow
to change the setpoint.
5
Release
SET
button. The new setting is saved.
6
Press and hold
Up
and
Down Arrows
simultaneously for 3 seconds to exit program mode.
13.6.3 Hysteresis
►
Allowable temperature variance on each side of the freezer setpoint.
NOTE
Hysteresis is factory-preset and should not be changed. Contact Helmer Technical
Service for instructions regarding changing the Hysteresis value.