Horizon Series™ Models
14 Maintenance
NOTE
►
Refer to the operation manual for the preventive maintenance schedule.
►
Before performing maintenance, protect items in freezer from extended exposure to
adverse temperature.
►
Allow freezer temperature to stabilize at setpoint after performing service or after
extended door opening.
14.1
Recharge Refrigerant
!
CAUTION
►
Review all safety instructions prior to recharging refrigerant. Refer to
Section I, Item
2 (Safety).
►
Maintenance should only be performed by trained refrigeration technicians.
NOTE
Use only non-CFC R-404A refrigerant.
Full initial refrigerant charge varies by model and power requirements, which can be found on the product
specification label.
Model
Power
Requirements
Initial Charge
105 model
115 V
11.0 oz (312 g)
230 V
18.5 oz (524 g)
Obtain:
►
Refrigerant
►
Calibrated pressure gauge (0 psi to 220 psi (0 kPa to 1520 kPa))
Add refrigerant:
1
Attach pressure gauge to the fittings on the refrigeration lines.
2
Monitor the low side (suction) pressure through a full compressor cycle.
3
Measure the pressure at the end of the next cycle, immediately before the compressor stops.
NOTE
Pressure varies depending on ambient air temperature.
4
Add refrigerant. Check the pressure on the high side and the low side.
►
Low side = 5 psi to 7 psi (34 kPa to 48 kPa)
►
High side = 180 psi to 220 psi (1241 kPa to 1517 kPa)
5
Remove pressure gauge.
14.2
Check Monitoring System Battery
The monitoring system does not indicate the charge level of the battery. Regularly test the battery.
Replace battery if the test fails or if the battery has been in use for one year.
Test the battery:
1
Switch the AC ON/OFF switch OFF.
a
Display should continue to display information.
b
If the display is blank, replace battery.
2
Switch AC ON/OFF switch ON.
NOTE
Use a battery which meets manufacturer’s specifications.