360142-D/D
77
Horizon Series™ Models
14.3
Check Optional Access Control System Battery
During an AC power failure, the Access Control backup battery provides backup power to power the
magnetic Access Control lock. Test the Access Control backup battery to ensure it is working properly.
Check the battery:
1
Ensure Access Control battery key switch is switched ON.
2
Switch AC ON/OFF switch OFF.
3
Verify “PoFF” (power failure) message is displayed.
4
Attempt to open the cabinet door.
►
If the door remains locked, the battery is functional.
►
If the door does not remain locked, replace the battery.
4
Switch AC ON/OFF switch ON.
14.4
Clean the Freezer
14.4.1
Condenser Grill
!
CAUTION
Disconnect freezer from AC power when cleaning condenser grill.
In environments where freezer is exposed to excessive lint or dust, condenser grill may require cleaning
more frequently than stated in preventive maintenance schedule.
Clean the condenser grill using a soft brush and a vacuum cleaner.
14.4.2 Exterior
Clean exterior surfaces with soft cotton cloth and non-abrasive liquid cleaner.
14.4.3 Interior
Clean painted surfaces with mild detergent. Clean stainless steel surfaces with a general-purpose
laboratory cleaner suitable for stainless steel.
14.4.4
Clean and Refill Probe Bottle
Obtain:
►
Fresh water-bleach solution (not provided)
►
1:9 ratio of bleach to water
►
Bleach is 5% solution of commercial sodium hypochlorite (NaOCl)
►
Equivalent oxidizing cleaner/disinfectant approved by your organization may be substituted
►
4 oz (120 mL) of product simulation solution per bottle
►
Solution is a 1:1 ratio of water to propylene glycol (or equivalent low-temperature fluid)
Clean and refill bottle:
1
Remove probe from bottle.
2
Remove bottle from bracket.
3
Clean bottle with water-bleach solution.
4
Fill bottle with 4 oz (120 mL) of product simulation solution.
5
Cap bottle tightly to minimize evaporation.
6
Place bottle in bracket.
7