360142-D/D
7
General Information
2 Safety
Includes general safety information for freezer operation.
2.1 Labels
Caution: Risk of damage
to equipment or danger to
operator
Caution: Unlock all casters
Caution: Hot surface
Earth / ground terminal
Caution: Shock/electrical
hazard
Protective earth / ground
terminal
2.2
Avoiding Injury
►
Review safety instructions before installing, using, or maintaining the equipment.
►
Do not open multiple, loaded drawers or baskets at the same time.
►
Before moving unit, ensure casters are unlocked and free of debris.
►
Never physically restrict any moving component.
►
Avoid removing electrical service panels and access panels unless so instructed.
►
Use manufacturer supplied power cords only.
!
CAUTION
Decontaminate parts prior to sending for service or repair. Contact Helmer or your
distributor for decontamination instructions and a Return Authorization Number.
3
Configuration
3.1
Model and Input Power
NOTE
Service information varies depending on the model and power requirements.
This information appears on the product specification label, located on the rear of the freezer. The model
also appears on a label located in the chamber on the upper side of the right wall.
REF
SN
ILF105
i.Series®
2000000
Voltage 115 V
HZ
60
Amps
5.75 A
Power 0.64 kW
Refrigerant Type: 404A
Made in USA
Version D
Laboratory Freezer
Weight 197 lb / 89 kg
14400 Bergen Boulevard
Noblesville, IN USA
www.helmerinc.com
2012
0086
UL 61010-1/CSA 61010-1
Certified
C
US
REF
SN
ILF105
2000000
Version D
www.helmerinc.com
D
A
B
C
Left: Chamber label. Right: Product Specification label (located on the rear at lower left).
Label
Description
A
Model (REF)
B
Serial number (SN)
C
Version
D
Power requirements