360137-D/D
10
General Information
6.3
Chart Recorder
A B F
C
D
E
Chart recorder with paper and battery installed.
Label Description
Function
A
Left and Right Arrow
buttons
Adjust settings and stylus position
B
LED
Indicates status of chart recorder in operating mode, or selected
temperature range in paper change mode
C
Chart change button
Adjust position of stylus when changing chart paper, or run a test pattern
D
Stylus
Mark temperature line on paper
E
Reset button
Restart chart recorder
F
Backup battery
Provides power during AC power failure. Connect prior to use.
6.3.1
Install and Change Chart Paper
1
Press and hold
C
button. When stylus begins to move left, release button. The LED flashes to
indicate current temperature range.
2
When stylus stops moving, remove chart knob then move knob up and away from chart paper.
3
Place new chart paper on chart recorder.
4
Gently lift stylus and rotate paper so current time line corresponds to time line groove.
5
Hold chart paper and reinstall chart knob.
NOTE
For accurate temperature reading, ensure that current time is aligned with time line
groove when chart knob is tightened.
6
Confirm the temperature range is set to the correct value.
7
Press and hold
C
button. When the stylus begins to move right, release the button.
8
Confirm the stylus is marking the temperature correctly.