360137-D/D
17
Horizon Series™ Models
9.5.1
Setting Parameter Values
1
Press and hold the
Up
and
Down Arrows
simultaneously for 3 seconds to enter program mode.
2
The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
3
Press and release
SEL
button to scroll through the parameters.
4
Once the desired parameter is selected, press and hold the
SET
button while pressing the
Up
or
Down Arrow
to select the desired value.
5
Release
SET
button. The new setting is saved.
6
Press and hold the
Up
and
Down Arrows
simultaneously for 3 seconds to exit program mode.
NOTE
Contact Helmer Technical Service for setting Hysteresis values.
9.5.2
Set Temperature Units
NOTE
If temperature units are changed, the temperature setpoints, offsets and alarm settings
must be recalibrated.
1 Press and hold the
Up
and
Down Arrows
simultaneously for 3 seconds to enter program mode.
2 The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
3 Press and hold the
SET
button while pressing the
Up
or
Down Arrow
to select the desired
temperature unit.
4 Release
SET
button. The new setting is saved.
5
Press and hold the
Up
and
Down Arrows
simultaneously for 3 seconds to exit program mode.
9.6
Temperature Calibration Offsets
Temperature calibration offsets indicate an acceptable margin of error between the actual temperature
value and the desired temperature value.
9.6.1
Monitor Offset
►
Value is factory-set to match an independent thermometer.
►
Refer to the service manual for instructions in changing the Monitor Offset.
9.6.2
Control Sensor Offset
►
Factory-set to match a calibrated reference thermometer.
►
Varies for each freezer.
NOTICE
Control Sensor Offset is factory-preset and should not be changed. Contact Helmer
Technical Service for instructions regarding changing the Control Sensor Offset.
9.6.3 Hysteresis
►
Allowable temperature variance on each side of the freezer setpoint.
NOTICE
Hysteresis is factory-preset and should not be changed. Contact Helmer Technical
Service for instructions regarding changing the Hysteresis value.