i.Series® Models
7.3
Set Alarm Parameters
>
> Alarm Settings
Control the conditions and timing of alarm condition indicators
displayed on the i.C³ Home screen. Touch
+
or
–
on spin box to set
each parameter.
7.4
Normal Operation
The i.C³ Home screen displays temperature and alarm information, and provides icons for reaching other
functions of the i.C³.
Home screen
Home screensaver (touch to return to Home screen)
7.5
Active Alarms
Home with no alarms.
Home with active alarm.
Alarm
Description
High Temperature
Chamber temperature reading is above high temperature alarm setpoint
Low Temperature
Chamber temperature reading is below low temperature alarm setpoint
Low Battery
Rechargeable battery voltage is low
Power Failure
Power to unit has been disrupted
Probe Failure
Probe not functioning properly
Door Open
Door is open beyond user-specified duration
Compressor
Temperature
Compressor temperature reading is above high temperature alarm setpoint
Communication
Failure Messages
1, 2, 3
1
Communication lost between i.C³ display board and control board
2
Communication lost between i.C³ display board and internal system memory
3
Corrupt database