360137-D/D
13
i.Series® Models
7.6
Mute and Disable Active Alarms
Audible alarms may be muted by touching the
Mute
button to set delay.
:15
Unmuted Muted
7.7
Access Control (Optional)
Allows user-specific secure access to the freezer.
NOTE
►
During a power failure, the backup battery provides power to the monitoring system
and the power failure alarm. If the backup battery is not functioning, the power failure
alarm will not be activated.
►
During a power failure, the backup battery continues to provide power to the optional
Access Control lock (if equipped). If the backup battery is not functioning, the
optional Access Control lock will not secure the door.
►
During a power failure, the optional Access Control lock will remain locked until
battery power is depleted or until the backup battery key switch is switched OFF.
►
Switching the backup battery key switch OFF will disable the monitoring system.
►
During a power failure, switch the battery backup switch OFF and use the mechanical
door key to provide secure storage for freezer contents.
7.7.1 Setup
>
>
> Access Setup
Configure and manage use-specific accounts to allow
controlled access to the freezer.
►
Enter the supervisor PIN to set up Access Control
►
Initial factory supervisor PIN = 5625
NOTE
The supervisor PIN can not be deleted, and should be changed to prevent unauthorized
user ID setup. The supervisor PIN does not allow access to the unit. At least one user
ID must be set up to gain access to the unit.