Horizon Series™ Models
2
View
maximum
temperature recording.
a
Press and hold the
Up Arrow
button for 1 second and listen for a single beep.
b
The display will alternate between
HI
and a valid temperature value five (5) times followed by a
single beep to indicate exit back to the temperature display.
3
View recorded temperature timer.
NOTE
►
The timer denotes the period of time that has elapsed. It does not display the time at
which a minimum or maximum temperature occurred.
►
The maximum period of time the timer can record is 99:59 (99 hours and 59 minutes).
a
Press and hold either the
Up
or
Down Arrow
button for 1 second.
b
While the display is flashing the
HI
or
LO
value, press and hold the
SET
button for 1 second.
c
The display will alternate five (5) times between
CLr
and a value representing the number of
hours and minutes that have elapsed since the last recording (example: 12:47 would represent 12
hours and 47 minutes). A single beep will follow to indicate exit back to temperature display.
4
Clear minimum and maximum temperature recordings.
a
Press and hold either the
Up
or
Down Arrow
button for 1 second.
b
While the display is flashing the
HI
or
LO
value, press and hold the
SET
button for 1 second and
listen for a single beep.
c
While the display is flashing the elapsed time since last reset, press and hold the
SET
button
for 2 seconds.
CLr
will be displayed followed by a series of 3 beeps to indicate exit back to the
temperature display.
NOTE
The minimum and maximum temperature and timer will reset when:
►
the unit is powered off and battery backup is not engaged, or
►
after 99 hours and 59 minutes have elapsed.
9.4
Change Freezer Temperature Setpoint
NOTE
Default setpoint is -30.0 °C
1
Press and release
SEL
to change to Control mode. The CONTROL lamp will illuminate.
2
Press and hold
SET
to display the current setpoint temperature.
3
Hold
SET
and press the
Up
or
Down Arrow
as necessary to set the desired setpoint value.
4
Release all buttons; the setpoint is changed.
5
Press and release
SEL
to return to Monitor mode. The MONITOR lamp will illuminate.
9.5
Table of Parameters
Parameter
Visual Indicator
Range
Default
Celsius or Fahrenheit
None
.C, .F
.C
High Temperature
MONITOR Lamp &
HIGH Lamp
-40.0 to 40.0 (°C)
-40 to 104 (°F)
-20.0°C
Low Temperature
MONITOR Lamp &
LOW Lamp
-40.0 to 40.0 (°C)
-40 to 104 (°F)
-40.0°C
Monitor Offset
MONITOR Lamp
-10.0 to 10.0 (°C)
-18 to 18 (°F)
Varies
Control Offset
CONTROL Lamp
-10.0 to 10.0 (°C)
-18 to 18 (°F)
Varies
Hysteresis
CONTROL Lamp
0.5 to 2.5 (°C)
1 to 5 (°F)
2.0°C