360137-D/D
19
Horizon Series™ Models
9.9
Access Control (Optional)
Allows user-specific secure access to the freezer.
NOTE
►
During a power failure, the backup battery continues to provide power to the optional
Access Control lock (if equipped). If the backup battery is not functioning, the
optional Access Control lock will not secure the door.
►
During a power failure, the optional Access Control lock will remain locked until
battery power is depleted or until the backup battery key switch is switched OFF.
►
During a power failure, switch the battery backup switch OFF and use the mechanical
door key to provide secure storage for freezer contents.
9.9.1 Setup
The Access Control keypad was programmed at the factory with a master code (0000). The master code
is used to program the keypad and enter user codes. The master code also releases the door lock.
NOTE
The master code can not be deleted, and should be changed to prevent unauthorized
user code setup.
Enter unique user codes for up to 100 users. Each user code is stored with a specific record location
number. Keep a log of the location numbers and user codes with users’ names.
9.9.2
Add User Code
►
Enter the master code
►
Press
1
to initiate user code programming function
►
Enter the location number (00 - 99)
►
Enter the user code (4 - 9 digit number)
►
Press
*
to save changes and return to normal operation
9.9.3
Delete User Code
►
Enter the master code
►
Press
1
to initiate user code programming function
►
Enter the location number (00 - 99)
►
Press
*
to save changes
9.9.4
Open Freezer with Access Control
3
2
1
6
5
4
9
8
7
#
0
►
Enter the user code
►
Press #