360142-D/D
59
Horizon Series™ Models
12.2.1
Set Current Time
1
Switch the AC ON/OFF switch OFF. Turn the Alarm Disable key switch OFF.
2
Remove the kick plate and pull out the electrical tray.
3
Rotate the time adjustment ring clockwise until the current time indicators show the current time.
4
Push the electrical tray in and replace the kick plate.
5
Switch the AC ON/OFF switch ON. Turn the Alarm Disable key switch ON.
6
Press the
Mute
button to disable the high temperature alarm while freezer reaches operating
temperature.
12.2.2
Set Defrost Time
►
Default setting of three defrost events per day
►
Default times are 12:00 AM, 8:00 AM, 4:00 PM
►
Defrost events can be added or removed
►
Defrost events can be scheduled for any time of day (in 15-minute intervals)
►
Defrost cycle lasts 15 to 30 minutes
►
Defrost events must be at least one hour apart
NOTE
Three defrost cycles are recommended for consistent freezer operation.
12.2.3
Schedule a Defrost Event
►
Switch defrost event switch(es) to the ON position (outside) to initiate a defrost event at that time.
►
Switch defrost event switch(es) to the OFF position (inside) to cancel a defrost event at that time.
1
Switch the AC ON/OFF switch OFF. Turn the Alarm Disable key switch OFF.
2
Remove the kick plate and pull out the electrical tray.
3
Add a defrost event:
►
At the appropriate time mark, position the switch ON (toward the outer ring).
4
Remove a defrost event:
►
At the appropriate time mark, position the switch OFF (toward the inside ring).
5
Push the electrical tray in and replace the kick plate.
6
Switch the AC ON/OFF switch ON. Turn the Alarm Disable key switch ON.
7
Press the
Mute
button to disable the high temperature alarm while freezer reaches operating
temperature.
Defrost controller with defrost times set to On position at 12:00 AM, 8:00 AM, and 4:00 PM.