360142-D/D
72
Horizon Series™ Models
13.3
Freezer Setpoint
NOTE
Default setpoint is -30.0 °C
Change the setpoint if:
►
Your organization requires a chamber temperature other than -30.0 °C.
Change setpoint.
1
On the monitoring system, press and release
SEL
to change to Control mode. The CONTROL lamp
will illuminate.
3
Press and hold
SET
to display the current temperature setpoint.
NOTE
The current temperature setpoint is typically higher than the chamber temperature.
4
Hold
SET
and press
Up Arrow
or
Down Arrow
to set the adjustment value.
5
Release
SET
button; the setpoint is changed.
6
Press and release
SEL
to return to Monitor mode. The MONITOR lamp will illuminate.
13.4
Table of Parameters
Parameter
Visual Indicator
Range
Default
Celsius or Fahrenheit
None
.C, .F
.C
High Temperature
MONITOR Lamp &
HIGH Lamp
-40.0 to 40.0 (°C)
-40 to 104 (°F)
-20.0°C
Low Temperature
MONITOR Lamp &
LOW Lamp
-40.0 to 40.0 (°C)
-40 to 104 (°F)
-40.0°C
Monitor Offset
MONITOR Lamp
-10.0 to 10.0 (°C)
-18 to 18 (°F)
Varies
Control Offset
CONTROL Lamp
-10.0 to 10.0 (°C)
-18 to 18 (°F)
Varies
Hysteresis
CONTROL Lamp
0.5 to 2.5 (°C)
1 to 5 (°F)
2.0°C
13.4.1
View Alarm Setpoints and Offset Values
1
Press and hold the
Up
and
Down Arrows
simultaneously for 3 seconds to enter program mode.
2
The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
3
Press and release
SEL
to scroll through the parameters and view settings.
4
Hold
Up
and
Down Arrows
simultaneously for 3 seconds to exit program mode.
5
The LED Display will show current monitor temperature.
13.4.2
Temperature Units
NOTE
If temperature units are changed, the temperature setpoints, offsets and alarm settings
must be recalibrated.
1
Press and hold the
Up
and
Down Arrows
simultaneously for 3 seconds to enter program mode.
2
The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
3
Press and hold the
SET
button while pressing the
Up
or
Down Arrow
to select the desired
temperature unit parameter.
4
Release
SET
button. The new setting is saved.
5
Press and hold
Up
and
Down Arrows
simultaneously for 3 seconds to exit program mode.