Helmer Scientific Platelet Agitator

Service Manual 

360391/A

11

3 Maintenance

Maintenance tasks should be completed according to the schedule below.

 Notes

•  These are recommended minimum requirements. Regulations for your organization or physical conditions at your  
  organization may require maintenance items to be performed more frequently, or only by designated service personnel.
•  The Preventive Maintenance schedule provided below supersedes the schedule provided in the Platelet Agitator Pro  

  Operation Manual.

Table 4. Preventive Maintenance Schedule

Task

Frequency

Annually

As Needed

Test the motion alarm.

ü

Check the 9V NiMH back-up battery for the motion alarm system after an 

extended power failure and change it if necessary, or change the battery if it has 
been in service for one year.

ü

Check moving parts for wear. Clean moving parts.

ü

Replace moving parts if worn.

ü

Clean the exterior of the platelet agitator.

ü

 Notes

• 

During a power failure (when the motion alarm is enabled), the back-up battery provides power to the motion alarm. 

 

  If the back-up battery is not functioning, the motion alarm will not be activated.
•  If the back-up battery does not provide power to the alarm system during the motion alarm test, replace the battery.
•  If battery has been in service for one year, replace battery.

• 

Use only manufacturer specified rechargeable battery for replacement (9V NiMH battery with a capacity of ≥175 mAh)

.

3.1 

Remove / Replace Trolley

The storage trolley must be removed when performing certain maintenance tasks.

 Remove Trolley

1. 

Switch the Motion Alarm ON/OFF switch 

OFF

.

2. 

Switch the Power ON/OFF switch 

OFF

, and disconnect the power cable from the power receptacle.

3. 

Remove the thumb screws securing the drawer stop panels to the left and right sides of the agitator trolley. (Note the 

 

  orientation of each panel)

4.  Carefully pull each panel from the agitator and set panels and thumb screws aside.
5. 

Slide the drawer(s) out and remove.

6. 

Use a #2 Phillips screwdriver to loosen the four (4) center screws on the floor of the trolley. 

7.  Slide the screw heads to the opposite end of the slots.
8.  Lift the trolley straight up and set aside.

 Install Trolley

1.  Carefully place the trolley on the base aligning the back and sides of the trolley with the back and sides of the base.  
 

Ensure the motor arm roller is placed within the trolley plastic guide block.

2. 

Slide the four (4) center screw heads in the floor of the trolley to the opposite ends of the slots, and secure using a 

 

 

#2 Phillips screwdriver.

3.  Reinstall drawers in desired location by aligning the outer edges of the drawer with the slots in the drawer guides and  
 

push inward.

4. 

Reinstall the drawer(s) stop panels in the same orientation as removed, and secure with thumb screws ensuring they are 

 

  finger tight.

5. 

Reconnect power cable to power receptacle and, switch the Power ON/OFF switch 

ON

.

6. 

Switch the Motion Alarm ON/OFF switch 

ON

.

Summary of Contents for i.Series PF15-Pro

Page 1: ...360391 A Platelet Agitator Service Manual i Series Pro Models Countertop PF15 Pro PF48 Pro PF96 Pro...

Page 2: ...19 Helmer Inc All other trademarks and registered trademarks are the property of their respective owners Helmer Inc doing business as DBA Helmer Scientific and Helmer Disclaimer This manual is intende...

Page 3: ...ply 6 2 3 Mounting Brackets 6 2 4 Placement Leveling and Setup 7 2 5 Configure Platelet Agitator for Use in a Platelet Incubator Optional 7 2 6 Storage Configuration 8 2 7 External Monitoring Devices...

Page 4: ...nterference at his own expense Helmer platelet agitators are intended to provide controlled agitation required for the storage of platelet products The devices are intended to be operated by personnel...

Page 5: ...on Never physically restrict any moving component Avoid removing electrical service panels and access panels unless so instructed Keep hands away from pinch points when agitation motion is enabled Avo...

Page 6: ...Note Service information varies depending on the model and power requirements This information appears on the product specification label on the back of the platelet agitator C A B Sample Product Spec...

Page 7: ...r Supply 1 Remove the cover plate by using the thumb to push and hold the spring loaded locking key downward while sliding the plate forward Retain the cover plate in secure location for future use 2...

Page 8: ...irect connection from the platelet agitator s alarm system to the platelet incubator s alarm system Platelet Agitator and Platelet Incubator Compatibility Note Only Pro Series platelet incubators can...

Page 9: ...tion Drawers can be removed or moved to adjust storage space as needed CAUTION To avoid injury and prevent damage to the agitator ensure both left and right side drawer stop panels are fully installed...

Page 10: ...AC RMS or 70V DC is connected to the J2 remote alarm circuit the remote alarm will not function properly may be damaged or may result in injury to the user J3 Agitator Powered Remote Alarm Interface T...

Page 11: ...me Alarms are sent through the remote alarm interface The table below indicates if an alarm is audible A visual V or sent through the remote alarm interface R Notes If the motion alarm switch is ON wi...

Page 12: ...ry for replacement 9V NiMH battery with a capacity of 175 mAh 3 1 Remove Replace Trolley The storage trolley must be removed when performing certain maintenance tasks Remove Trolley 1 Switch the Motio...

Page 13: ...Alarm Volume is set to an audible level and the Alarm Delay is set to the minimum delay 2 Ensure the Motion Alarm ON OFF switch is ON 3 Switch the Power ON OFF switch OFF to stop agitation When the a...

Page 14: ...d turn on the power If operation is noisy or jerky replace the gearbox and motor Agitator is moving around on placement surface Placement surface is slick Relocate agitator to non slick surface Instal...

Page 15: ...C Drawer white 15 322731 1 069 J Motion Alarm Time Delay Switch 48 322749 1 069 K Agitator Communication Switch 96 322756 1 069 L Data Port for use with Pro Series Incubator Drawer stainless steel 15...

Page 16: ...ription Model Part Number A Wheel includes 4 wheels and brackets 801031 1 E Motor 15 48 801057 1 B Actuator Sensor 801035 1 96 801032 1 C 9V NiMH Rechargeable Back up Battery 120232 F Crank Block 8010...

Page 17: ...Helmer Scientific Platelet Agitator Service Manual 360391 A 16 Appendix A Wiring Diagram...

Page 18: ...ed at a location other than Helmer or for repairs made without the assistance of a Helmer technician will be the responsibility of the end user Additional Warranty Information The time periods set for...

Page 19: ...Helmer Scientific Platelet Agitator Service Manual 360391 A 18...

Page 20: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA Copyright 2019 Helmer Inc 360391 A...

Reviews: