Helmer Scientific Platelet Agitator
Service Manual
360391/A
12
3.2
Motion Alarm Test
Test the motion alarm to ensure it activates when the platelet agitator motion stops. The alarm may not activate immediately,
depending on the motion alarm delay setting. If the alarm does not activate when it should, refer to
Troubleshooting.
Notes
• The motion alarm delay range is 10 - 600 seconds. When testing the alarm, be certain the set delay time has passed.
•
When the motion alarm is activated, the red LED on the motion alarm switch will flash.
• The motion alarm is muted when the alarm volume is turned fully counterclockwise.
Test Motion Alarm
1.
Ensure the Alarm Volume is set to an audible level and the Alarm Delay is set to the
minimum delay (as noted by red rotational arrows).
2.
Switch the Power ON/OFF switch
OFF
to stop agitation. When the alarm delay
period elapses, the motion alarm will activate.
3.
Switch the Power ON/OFF switch
ON
.
3.3
Test and Replace Motion Alarm Back-up Battery
The motion alarm system is powered by DC power, backed up by a rechargeable 9V NiMH battery. The battery charges when the
platelet agitator is connected to its external AC/DC power supply. In the event of a facility AC power failure, agitation would stop
and the battery will power the motion alarm.
Test Motion Alarm Back-up Battery
1.
Ensure the Alarm Volume is set to an audible level and the Alarm Delay is set to the minimum delay.
2.
Ensure the Motion Alarm ON/OFF switch is
ON
.
3.
Switch the Power ON/OFF switch
OFF
to stop agitation. When the alarm delay period elapses, the motion alarm
will activate. If the motion alarm does not activate, replace the battery.
4.
Switch the Power ON/OFF switch
ON
.
CAUTION
To prevent personal injury or damage to the agitator, use only a 9V NiMH rechargeable battery.
Replace Motion Alarm Back-up Battery
1.
Switch the Motion Alarm ON/OFF switch
OFF
.
2.
Switch the Power ON/OFF switch
OFF
.
3. Disconnect the DC power cord from the rear power jack.
4. Remove the trolley to access the back-up battery in the base
of the unit.
5. Disconnect the battery from the terminals and remove.
6. Insert the new battery and connect to the terminals ensuring the new
battery is fully seated in the holder.
7. Reinstall the trolley.
8. Reconnect the DC power cord to the power receptacle.
9.
Switch the Power ON/OFF switch
ON
.
10.
Switch the Motion Alarm ON/OFF switch
ON
.
3.4
Clean Agitator
Clean exterior surfaces with a soft cotton cloth and non-abrasive liquid cleaner. Clean shelves and drawer surfaces with a soft
brush. Disinfect using a mild disinfectant solution.
The interior base of the agitator may be cleaned using a clean compressed air cleaner to blow out any accumulated dust, or wiped
out with a soft cotton cloth.
Note
Helmer recommends cleaning the interior base only when inspecting or replacing moving parts or battery.
Alarm Volume
Alarm Delay