Helmer Scientific Platelet Agitator

Service Manual 

360391/A

12

3.2 

Motion Alarm Test

Test the motion alarm to ensure it activates when the platelet agitator motion stops. The alarm may not activate immediately, 
depending on the motion alarm delay setting. If the alarm does not activate when it should, refer to 

Troubleshooting.

 Notes

•  The motion alarm delay range is 10 - 600 seconds. When testing the alarm, be certain the set delay time has passed.
• 

When the motion alarm is activated, the red LED on the motion alarm switch will flash.

•  The motion alarm is muted when the alarm volume is turned fully counterclockwise.

 Test Motion Alarm

1. 

Ensure the Alarm Volume is set to an audible level and the Alarm Delay is set to the 

 

  minimum delay (as noted by red rotational arrows).

2. 

Switch the Power ON/OFF switch 

OFF

 to stop agitation. When the alarm delay 

 

period elapses, the motion alarm will activate. 

3. 

Switch the Power ON/OFF switch 

ON

3.3 

Test and Replace Motion Alarm Back-up Battery

The motion alarm system is powered by DC power, backed up by a rechargeable 9V NiMH battery. The battery charges when the 

platelet agitator is connected to its external AC/DC power supply. In the event of a facility AC power failure, agitation would stop 
and the battery will power the motion alarm.

 Test Motion Alarm Back-up Battery

1. 

Ensure the Alarm Volume is set to an audible level and the Alarm Delay is set to the minimum delay. 

2. 

Ensure the Motion Alarm ON/OFF switch is 

ON

.

3. 

Switch the Power ON/OFF switch 

OFF

 to stop agitation. When the alarm delay period elapses, the motion alarm  

 

will activate. If the motion alarm does not activate, replace the battery.

4. 

Switch the Power ON/OFF switch 

ON

.

CAUTION

To prevent personal injury or damage to the agitator, use only a 9V NiMH rechargeable battery.

 Replace Motion Alarm Back-up Battery

1. 

Switch the Motion Alarm ON/OFF switch 

OFF

.

2. 

Switch the Power ON/OFF switch 

OFF

.

3.  Disconnect the DC power cord from the rear power jack.
4.  Remove the trolley to access the back-up battery in the base  
 

of the unit.

5.  Disconnect the battery from the terminals and remove.
6.  Insert the new battery and connect to the terminals ensuring the new  
 

battery is fully seated in the holder.

7.  Reinstall the trolley.
8.  Reconnect the DC power cord to the power receptacle.
9. 

Switch the Power ON/OFF switch 

ON

.

10. 

Switch the Motion Alarm ON/OFF switch 

ON

.

3.4 

Clean Agitator

Clean exterior surfaces with a soft cotton cloth and non-abrasive liquid cleaner. Clean shelves and drawer surfaces with a soft 
brush. Disinfect using a mild disinfectant solution.

The interior base of the agitator may be cleaned using a clean compressed air cleaner to blow out any accumulated dust, or wiped 
out with a soft cotton cloth.

 Note

Helmer recommends cleaning the interior base only when inspecting or replacing moving parts or battery. 

Alarm Volume

Alarm Delay

Summary of Contents for i.Series PF15-Pro

Page 1: ...360391 A Platelet Agitator Service Manual i Series Pro Models Countertop PF15 Pro PF48 Pro PF96 Pro...

Page 2: ...19 Helmer Inc All other trademarks and registered trademarks are the property of their respective owners Helmer Inc doing business as DBA Helmer Scientific and Helmer Disclaimer This manual is intende...

Page 3: ...ply 6 2 3 Mounting Brackets 6 2 4 Placement Leveling and Setup 7 2 5 Configure Platelet Agitator for Use in a Platelet Incubator Optional 7 2 6 Storage Configuration 8 2 7 External Monitoring Devices...

Page 4: ...nterference at his own expense Helmer platelet agitators are intended to provide controlled agitation required for the storage of platelet products The devices are intended to be operated by personnel...

Page 5: ...on Never physically restrict any moving component Avoid removing electrical service panels and access panels unless so instructed Keep hands away from pinch points when agitation motion is enabled Avo...

Page 6: ...Note Service information varies depending on the model and power requirements This information appears on the product specification label on the back of the platelet agitator C A B Sample Product Spec...

Page 7: ...r Supply 1 Remove the cover plate by using the thumb to push and hold the spring loaded locking key downward while sliding the plate forward Retain the cover plate in secure location for future use 2...

Page 8: ...irect connection from the platelet agitator s alarm system to the platelet incubator s alarm system Platelet Agitator and Platelet Incubator Compatibility Note Only Pro Series platelet incubators can...

Page 9: ...tion Drawers can be removed or moved to adjust storage space as needed CAUTION To avoid injury and prevent damage to the agitator ensure both left and right side drawer stop panels are fully installed...

Page 10: ...AC RMS or 70V DC is connected to the J2 remote alarm circuit the remote alarm will not function properly may be damaged or may result in injury to the user J3 Agitator Powered Remote Alarm Interface T...

Page 11: ...me Alarms are sent through the remote alarm interface The table below indicates if an alarm is audible A visual V or sent through the remote alarm interface R Notes If the motion alarm switch is ON wi...

Page 12: ...ry for replacement 9V NiMH battery with a capacity of 175 mAh 3 1 Remove Replace Trolley The storage trolley must be removed when performing certain maintenance tasks Remove Trolley 1 Switch the Motio...

Page 13: ...Alarm Volume is set to an audible level and the Alarm Delay is set to the minimum delay 2 Ensure the Motion Alarm ON OFF switch is ON 3 Switch the Power ON OFF switch OFF to stop agitation When the a...

Page 14: ...d turn on the power If operation is noisy or jerky replace the gearbox and motor Agitator is moving around on placement surface Placement surface is slick Relocate agitator to non slick surface Instal...

Page 15: ...C Drawer white 15 322731 1 069 J Motion Alarm Time Delay Switch 48 322749 1 069 K Agitator Communication Switch 96 322756 1 069 L Data Port for use with Pro Series Incubator Drawer stainless steel 15...

Page 16: ...ription Model Part Number A Wheel includes 4 wheels and brackets 801031 1 E Motor 15 48 801057 1 B Actuator Sensor 801035 1 96 801032 1 C 9V NiMH Rechargeable Back up Battery 120232 F Crank Block 8010...

Page 17: ...Helmer Scientific Platelet Agitator Service Manual 360391 A 16 Appendix A Wiring Diagram...

Page 18: ...ed at a location other than Helmer or for repairs made without the assistance of a Helmer technician will be the responsibility of the end user Additional Warranty Information The time periods set for...

Page 19: ...Helmer Scientific Platelet Agitator Service Manual 360391 A 18...

Page 20: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA Copyright 2019 Helmer Inc 360391 A...

Reviews: