Helmer Scientific Platelet Agitator

Service Manual 

360391/A

13

4 Troubleshooting

CAUTION

Review all safety instructions prior to troubleshooting.

4.1 

General Operation Problems

Problem

Possible Cause

Action

Agitator is inoperable

Power switch is in OFF position.

Switch power switch to ON.

The power cord is disconnected or 
loose.

Ensure cord is properly attached.

If the platelet agitator is installed in 
a Helmer incubator, the door switch 
has not been activated to allow 
agitation to resume.

Close the incubator door. Verify i.C

3

 AgiTrak screen door status is closed.

Agitation speed will not 
change.

Stand-alone agitator 
communication switch is not set to 

the (X) position

Verify agitator communication switch is set to the (X) position for stand-

alone mode.

Agitation is noisy or rough.

Trolley is not evenly loaded.

Verify product drawers are evenly loaded.

Insufficient clearance around the 

trolley obstructing agitation.

Verify agitator has proper clearance, and no objects are preventing 

agitation.

One or more wheels are worn.

Inspect the wheels. Replace if necessary.

Motor or gearbox is faulty.

With the gearbox and motor exposed, turn on the power. If operation is 
noisy or jerky, replace the gearbox and motor. 

Agitator is moving around on 
placement surface.

Placement surface is slick.

Relocate agitator to non-slick surface.

Install mounting brackets (included in packaging) and attach to surface.

Agitation speed is set too high.

Reduce agitation speed.

4.2 

Alarm Activation Problems

Problem

Possible Cause

Action

Agitation has stopped, but 
the motion alarm is not 
active. Stand-alone, not in 
incubator.

The agitation alarm selector switch 
is in wrong position.

Verify agitator communication switch position is (X) for standalone mode.

Alarm delay period has not lapsed

Check alarm delay setting.

The agitator motion alarm switch 
is 

OFF

.

Turn the agitator motion alarm switch 

ON

.

The alarm activates while the 
trolley is moving. 

(stand-alone)

The motion actuator and motion 
sensor are not aligned.

Ensure the actuator on the bottom of the trolley is aligned with the sensor 
on the base. Adjust the position if not aligned.

The motion sensor or actuator is 
faulty.

Check the connections for the actuator and the sensor. Replace the 
actuator or the sensor if faulty.

The Agitator Control Board is faulty.

Replace the Agitator Control Board.

Alarm switch is blinking, but 
no audible alarm is active.

Alarm volume is muted.

Rotate alarm volume clockwise to increase alarm volume.

Summary of Contents for i.Series PF15-Pro

Page 1: ...360391 A Platelet Agitator Service Manual i Series Pro Models Countertop PF15 Pro PF48 Pro PF96 Pro...

Page 2: ...19 Helmer Inc All other trademarks and registered trademarks are the property of their respective owners Helmer Inc doing business as DBA Helmer Scientific and Helmer Disclaimer This manual is intende...

Page 3: ...ply 6 2 3 Mounting Brackets 6 2 4 Placement Leveling and Setup 7 2 5 Configure Platelet Agitator for Use in a Platelet Incubator Optional 7 2 6 Storage Configuration 8 2 7 External Monitoring Devices...

Page 4: ...nterference at his own expense Helmer platelet agitators are intended to provide controlled agitation required for the storage of platelet products The devices are intended to be operated by personnel...

Page 5: ...on Never physically restrict any moving component Avoid removing electrical service panels and access panels unless so instructed Keep hands away from pinch points when agitation motion is enabled Avo...

Page 6: ...Note Service information varies depending on the model and power requirements This information appears on the product specification label on the back of the platelet agitator C A B Sample Product Spec...

Page 7: ...r Supply 1 Remove the cover plate by using the thumb to push and hold the spring loaded locking key downward while sliding the plate forward Retain the cover plate in secure location for future use 2...

Page 8: ...irect connection from the platelet agitator s alarm system to the platelet incubator s alarm system Platelet Agitator and Platelet Incubator Compatibility Note Only Pro Series platelet incubators can...

Page 9: ...tion Drawers can be removed or moved to adjust storage space as needed CAUTION To avoid injury and prevent damage to the agitator ensure both left and right side drawer stop panels are fully installed...

Page 10: ...AC RMS or 70V DC is connected to the J2 remote alarm circuit the remote alarm will not function properly may be damaged or may result in injury to the user J3 Agitator Powered Remote Alarm Interface T...

Page 11: ...me Alarms are sent through the remote alarm interface The table below indicates if an alarm is audible A visual V or sent through the remote alarm interface R Notes If the motion alarm switch is ON wi...

Page 12: ...ry for replacement 9V NiMH battery with a capacity of 175 mAh 3 1 Remove Replace Trolley The storage trolley must be removed when performing certain maintenance tasks Remove Trolley 1 Switch the Motio...

Page 13: ...Alarm Volume is set to an audible level and the Alarm Delay is set to the minimum delay 2 Ensure the Motion Alarm ON OFF switch is ON 3 Switch the Power ON OFF switch OFF to stop agitation When the a...

Page 14: ...d turn on the power If operation is noisy or jerky replace the gearbox and motor Agitator is moving around on placement surface Placement surface is slick Relocate agitator to non slick surface Instal...

Page 15: ...C Drawer white 15 322731 1 069 J Motion Alarm Time Delay Switch 48 322749 1 069 K Agitator Communication Switch 96 322756 1 069 L Data Port for use with Pro Series Incubator Drawer stainless steel 15...

Page 16: ...ription Model Part Number A Wheel includes 4 wheels and brackets 801031 1 E Motor 15 48 801057 1 B Actuator Sensor 801035 1 96 801032 1 C 9V NiMH Rechargeable Back up Battery 120232 F Crank Block 8010...

Page 17: ...Helmer Scientific Platelet Agitator Service Manual 360391 A 16 Appendix A Wiring Diagram...

Page 18: ...ed at a location other than Helmer or for repairs made without the assistance of a Helmer technician will be the responsibility of the end user Additional Warranty Information The time periods set for...

Page 19: ...Helmer Scientific Platelet Agitator Service Manual 360391 A 18...

Page 20: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA Copyright 2019 Helmer Inc 360391 A...

Reviews: