Service Manual 

360391/A

8

 

    

PC900-Pro with  PF48-Pro installed. 

 

 

 

 

Agitator communication switch

Helmer Pro Series platelet agitators can be installed inside Helmer Pro Series platelet incubators. The agitator can be connected 
to the incubator through a data cable. The connection allows the platelet incubator to monitor the platelet agitator motion, and to 
sound an alarm when agitation motion stops. 
Power is supplied to the agitator through the coiled DC power cable. Connect the data cable and DC power cable supplied with the 
incubator prior to placing the agitator inside the incubator. The power cable has a rotating locking feature on each end to ensure a 

secure connection. When installing the agitator inside the incubator, finger tighten each rotating lock. The agitator communication 

switch must be set to position 1 (shown in red for visibility in the picture above) for use in the incubator.

2.6  Storage Configuration

Drawers can be removed or moved to adjust storage space as needed.

CAUTION

To avoid injury and prevent damage to the agitator, ensure both left and right side drawer stop panels are fully installed 
prior to operating the agitator.

 Remove and Replace Drawers

1.  Remove the thumb screws securing the drawer stop panels to the left and right sides of the  

  agitator. (Note the orientation of each panel)

2.  Carefully pull each panel from the agitator and set panels and thumb screws aside.
3. 

Slide the drawer(s) out and remove.

4. 

Reinstall drawer(s) in desired location by aligning the outer edges of the drawer with the slots 

 

 

in the drawer guides and push inward.

5.  Reinstall both drawer stop panels in the same orientation as removed, and secure with thumb  

  screws ensuring they are finger tight.

 

 

 

 Install Label Holders (optional)

1.  Insert the tabs on the label holder into the slots on the drawer.
2.  Pivot the holder around the drawer handle and align the hole on the label holder with the corresponding hole on  
 

the drawer.

3.  Push thumb screw through the hole in the label holder and through the hole in the drawer to secure.

Agitator Power Connection

Thumb Screws

Summary of Contents for i.Series PF15-Pro

Page 1: ...360391 A Platelet Agitator Service Manual i Series Pro Models Countertop PF15 Pro PF48 Pro PF96 Pro...

Page 2: ...19 Helmer Inc All other trademarks and registered trademarks are the property of their respective owners Helmer Inc doing business as DBA Helmer Scientific and Helmer Disclaimer This manual is intende...

Page 3: ...ply 6 2 3 Mounting Brackets 6 2 4 Placement Leveling and Setup 7 2 5 Configure Platelet Agitator for Use in a Platelet Incubator Optional 7 2 6 Storage Configuration 8 2 7 External Monitoring Devices...

Page 4: ...nterference at his own expense Helmer platelet agitators are intended to provide controlled agitation required for the storage of platelet products The devices are intended to be operated by personnel...

Page 5: ...on Never physically restrict any moving component Avoid removing electrical service panels and access panels unless so instructed Keep hands away from pinch points when agitation motion is enabled Avo...

Page 6: ...Note Service information varies depending on the model and power requirements This information appears on the product specification label on the back of the platelet agitator C A B Sample Product Spec...

Page 7: ...r Supply 1 Remove the cover plate by using the thumb to push and hold the spring loaded locking key downward while sliding the plate forward Retain the cover plate in secure location for future use 2...

Page 8: ...irect connection from the platelet agitator s alarm system to the platelet incubator s alarm system Platelet Agitator and Platelet Incubator Compatibility Note Only Pro Series platelet incubators can...

Page 9: ...tion Drawers can be removed or moved to adjust storage space as needed CAUTION To avoid injury and prevent damage to the agitator ensure both left and right side drawer stop panels are fully installed...

Page 10: ...AC RMS or 70V DC is connected to the J2 remote alarm circuit the remote alarm will not function properly may be damaged or may result in injury to the user J3 Agitator Powered Remote Alarm Interface T...

Page 11: ...me Alarms are sent through the remote alarm interface The table below indicates if an alarm is audible A visual V or sent through the remote alarm interface R Notes If the motion alarm switch is ON wi...

Page 12: ...ry for replacement 9V NiMH battery with a capacity of 175 mAh 3 1 Remove Replace Trolley The storage trolley must be removed when performing certain maintenance tasks Remove Trolley 1 Switch the Motio...

Page 13: ...Alarm Volume is set to an audible level and the Alarm Delay is set to the minimum delay 2 Ensure the Motion Alarm ON OFF switch is ON 3 Switch the Power ON OFF switch OFF to stop agitation When the a...

Page 14: ...d turn on the power If operation is noisy or jerky replace the gearbox and motor Agitator is moving around on placement surface Placement surface is slick Relocate agitator to non slick surface Instal...

Page 15: ...C Drawer white 15 322731 1 069 J Motion Alarm Time Delay Switch 48 322749 1 069 K Agitator Communication Switch 96 322756 1 069 L Data Port for use with Pro Series Incubator Drawer stainless steel 15...

Page 16: ...ription Model Part Number A Wheel includes 4 wheels and brackets 801031 1 E Motor 15 48 801057 1 B Actuator Sensor 801035 1 96 801032 1 C 9V NiMH Rechargeable Back up Battery 120232 F Crank Block 8010...

Page 17: ...Helmer Scientific Platelet Agitator Service Manual 360391 A 16 Appendix A Wiring Diagram...

Page 18: ...ed at a location other than Helmer or for repairs made without the assistance of a Helmer technician will be the responsibility of the end user Additional Warranty Information The time periods set for...

Page 19: ...Helmer Scientific Platelet Agitator Service Manual 360391 A 18...

Page 20: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA Copyright 2019 Helmer Inc 360391 A...

Reviews: