The monitor or sensor must be positioned
where it can ‘see’ the room pressure of the
laboratory. The back connection spigot of the
monitor or sensor will accept the 25mm (1”)
hose which should be connected to the inner
chamber of the fume hood. The ideal position
for the hose connection is 100mm (4”) back
from the sash and 100mm (4”) higher than the
normal sash opening height.
AFA500 or Remote Airfl ow Sensor Installation Diagram
Monitor Panel Dimensions
Monitor Panel Cutout Dimensions
Fume hood
inner skin
Fume hood
outer skin
Normal Sash
working height
TEL Airfl ow Monitor Model AFA 500 MK 3
Installation, Operation, Maintenance Manual