The monitor or sensor must be positioned 
where it can ‘see’ the room pressure of the 
laboratory. The back connection spigot of the 
monitor or sensor will accept the 25mm (1”) 
hose which should be connected to the inner 
chamber of the fume hood. The ideal position 
for the hose connection is 100mm (4”) back 
from the sash and 100mm (4”) higher than the 
normal sash opening height.

AFA500 or Remote Airfl ow Sensor Installation Diagram

Monitor Panel Dimensions

Monitor Panel Cutout Dimensions

Fume hood
inner skin

Fume hood
outer skin

Normal Sash
working height

 TEL Airfl ow Monitor Model AFA 500 MK 3

Installation, Operation, Maintenance Manual