CONTRAINDICATIONS

The  Hemostatix  Thermal  Scalpel  System  is  contraindicated  in  the 

presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR 

or OXYGEN or NITROUS OXIDE

.

POTENTIAL ADVERSE EFFECTS

Known potential adverse effects include, but are not limited to, 

thermal injury to tissue, including nerves or other delicate tissues, and 

inability to effectively provide hemostasis of larger vessels.

SYSTEM WARNINGS

1.

EXPLOSION HAZARD – The Hemostatix Thermal Scalpel

System is contraindicated in the presence of a FLAMMABLE

ANESTHETIC MIXTURE WITH AIR or OXYGEN or

NITROUS OXIDE.

2.

NO MODIFICATION OF THIS EQUIPMENT IS

 

ALLOWED.

3.

Electrical shock hazard.  Do not remove cover.  Refer to

manufacturer for service.

4.

Do not attach unapproved components to the HTSS unit to

 

avoid

electrical shock.

5. Ensure the sound volume is adequately adjusted to that alarms

and alerts are clearly heard

6.

To avoid risk of electric shock, this equipment must only be

connected to a supply mains with protective earth.

7.

Do not place the HTSS controller unit in direct contact with

 

or

within 1 m of any type of electro-surgical equipment. This

equipment has been tested and found to comply with the limits

 

for

medical devices to the IEC 60601-1-

2:2014

. These limits

 

are

designed to provide reasonable protection against harmful

interference in a typical medical installation. This equipment

radiates electromagnetic fields and, if not installed and used in

accordance with the instructions, may cause harmful interference.

However, there is no guarantee that interference will not occur

 

in

 

a

particular installation. If this equipment does cause harmful

interference to other devices, which can be determined by

 

turning

the equipment off and on, or is affected by interference

INTRODUCTION

HEMOSTATIX THERMAL SCALPEL BLADES are similar in size, 

shape, and sharpness to traditional steel scalpel blades; however, 

Hemostatix blades can be heated to a user-selected temperature 

appropriate for sealing small vessels as they are cut.
To cut and simultaneously seal blood vessels effectively with minimum 

tissue damage, the sharp steel cutting edge of a heated scalpel blade 

must be uniformly maintained at the desired temperature within 

narrow limits. The Hemostatix Thermal Scalpel System utilizes micro-

circuitry incorporated within the blade itself to maintain the cutting 

edge temperature within the necessary tolerance, selectively delivering 

additional thermal energy only to those regions of the blade using heat 

due to tissue contact. By so doing, the Hemostatix Thermal Scalpel 

System automatically compensates for the varying degrees of heat loss 

that occur during surgical procedures (depending on the type of tissue 

being incised and the rate at which cutting is carried out), maintaining 

the cutting edge in the desired temperature range.
In contrast with electrosurgical devices, the Hemostatix Thermal Scalpel 

System passes no electrical current through the patient, and there is no 

sparking or electrical arcing to the tissue. Electrosurgical devices “cut” 

and/or cauterize using electrical currents which pass through the patient 

vaporizing tissue at the point of contact and generating heat and tissue 

damage down the path of the electrical current. The Hemostatix 

Thermal Scalpel System cuts tissue with a sharpened steel edge, like a 

conventional cold-steel scalpel blade, and simultaneously seals blood 

vessels using heat thermally conducted to the tissue from an elevated-

temperature blade which is electrically insulated from the patient. By 

thermally transferring heat from 

a uniformly- controlled, essentially constant temperature blade, 

the amount of tissue damage associated with hemostatic cutting is 

minimized

.

INDICATIONS FOR USE

The Hemostatix Thermal Scalpel System is a surgical instrument 

designed to retain the precise, clean cutting characteristics of 

the traditional steel scalpel while minimizing blood loss by 

simultaneously sealing blood vessels as they are cut, with minimum 

damage to surrounding tissue and virtually no muscle stimulation, using 

heat thermally conducted to the tissue from an elevated-temperature 

blade.

Rx Only – CAUTION:  Federal (U.S.A.) law restricts this device to 

sale by or on the order of a physician.

NOTICE TO USERS AND/OR PATIENTS

Any serious incident that has occurred in relation to the device should 

be reported to the manufacturer and the competent authority of the 

Member State in which the user and/or patient is established

.

Summary of Contents for 7013-8400

Page 1: ...OPERATING MANUAL THERMAL SCALPEL SYSTEM MODEL PRECISION 8400 CONTROLLER Part Number 7013 8400 For use with the Model 9050 Disposable Scalpel Handles...

Page 2: ...pel blade must be uniformly maintained at the desired temperature within narrow limits The Hemostatix Thermal Scalpel System utilizes micro circuitry incorporated within the blade itself to maintain t...

Page 3: ...TSS before cleaning the unit to avoid electrical shock 5 To avoid the risk of electrical shock achieve electrical grounding reliability with proper connections Connect the HTSS unit to hospital grade...

Page 4: ...dal 25 GROUNDING Reliability can only be achieved when the equipment is connected to a properly equivalent receptacle marked Hospital Grade 26 STERILIZATION The Model P8400 Hemostatix Thermal Scalpel...

Page 5: ...into the handle as this would damage the blade s imprinted circuitry and render it inoperable 11 The handle must be energized for the COAG switch to work 12 DO NOT immerse the handle in liquid of ANY...

Page 6: ...s may become broken and the blade stop working 4 FOOTSWITCH An optional footswitch REF 7013 8410 is available which allows the surgeon to set the desired temperatures of the blades as well as activate...

Page 7: ...y packaged sterile and ready for use They are discarded when they become dull just like conventional cold steel scalpel blades Blades are single use only and should never be reused 2 Automatic Calibra...

Page 8: ...Thermal Scalpel Controller REF 7013 8400 is compatible with the following Hemostatix Thermal Scalpel handle cable configurations REF 7013 9050 Model P8400 Handle w Integral Cable Assembly Qty 1 REF 7...

Page 9: ...te the thermal COAG mode depress the blue Thermal COAG switch on the top of the handle The blade temperature will approximate 300 C in the thermal COAG mode NOTE The handle must be energized for the C...

Page 10: ...auxiliary output connector is to potentially offer other modules i e irrigation suction etc that would complement the existing Hemostatix technology WARNING The auxiliary port is not to be utilized fo...

Page 11: ...oagulate on the blade thermally insulating it 5 CHANGING THE BLADE If the blade becomes dull or the change blade message appears on the controller display switch the handle ON OFF switch to off and wa...

Page 12: ...E Before cleaning the controller detach the controller unit from the AC power source servicing The Model P8400 Hemostatix Thermal Scalpel System consists of the controller unit a handle and a blade If...

Page 13: ...tained before any part of the MODEL P8400 HTSS is returned NOTE Handles and blades being returned must be cleaned sterilized and packaged in sterile packaging with labeling which verifies the sterilit...

Page 14: ...statix Thermal Scalpel System Blades are available in the following configurations similar to cold stainless steel scalpel blades No 10 No 12 and No 15 blades Model P8400 Hemostatix Thermal Scalpel Sy...

Page 15: ...Manufacturer Manufacturing date Maximum DC Output Caution Authorized Representative in the European Community Temperature decrease Temperature increase Volume increase decrease Auxilliary output port...

Page 16: ...can affect the operation of the product GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS The Model P8400 Hemostatix Thermal Scalpel System is intended for use in the electromagnetic e...

Page 17: ...e user of the Model P8400 Hemostatix Thermal Scalpel System should assure that it is used in such an environment Immunity Test Test Level Compliance Level Electromagnetic Environmental Guidance Conduc...

Page 18: ...cording to frequency of transmitter m 150 kHz to 80 MHz outside ISM bands d 3 5 V1 P 80 kHz to 800 MHz in ISM bands d 3 5 E1 P 800 MHz to 2 5 GHz d 7 E1 P 0 01 d 3 5 3 0 01 d 3 5 3 0 01 d 7 3 0 01 0 1...

Page 19: ...TE 109 Bartlett TN 38133 USA Telephone 1 901 261 0012 Quality First International OU Laki 30 12915 Tallinn Estonia Hemostatix Medical Technologies LLC All rights reserved HEMOSTATIX is a U S A Registe...

Reviews: