temperature, and then reinsert it.
6.
Never use any type of abrasive pad to clean the blades. The
abrasives will damage the circuit and render the blade unusable.
7.
To remove a blade from the handle, pull the blade straight out of
the handle. Bending, twisting or flexing the blade could damage
the blade contacts and retainers within the handle causing it to no
longer function.
8.
If you are getting multiple error messages during blade insertion,
try inserting the blade into the handle BEFORE plugging the
handle into the controller unit.
9.
DO NOT allow saline, or any other fluid, to enter the handle
during use. Saline is highly conductive and will interfere with
the
circuitry inside the handle causing the handle not to work
properly.
10.
DO NOT use any type of instrument (e.g. hemostats) to insert
the blade into the handle as this would damage the blade’s
imprinted circuitry and render it inoperable.
11.
The handle must be energized for the COAG switch to work.
12.
DO NOT immerse the handle in liquid of ANY KIND. The
handle contains electronic contacts and moisture sensitive
electrical components which can be damaged and fail to function
if immersed in liquids of any kind. DO NOT allow any solution
to penetrate to the interior of the handle.
13.
External Cleaning is the only controller maintenance that can be
performed by the user.
14.
Servicing the controller unit by other than and qualified service
personnel approved by Hemostatix Medical Technologies, LLC
renders the Warranty void.
15. Before cleaning the controller, detach the controller unit from
the AC power source.
16. DO NOT immerse the console.
17. DO NOT use an abrasive cloth or cleaners, especially on the
display screen.
18. DO NOT dispose of this product in the unsorted municipal
waste stream. Dispose of this product according to local
regulations.