- 10 -
Output function
Distribution output:
support up to 3 groups of transmitter isolated
distri24VDC, each group of distribution ≤ 60mA, support other
specifications isolated distribution (such as 12VDC, 5VDC distribution
output)
Transmission output:
support up to 4 channels of standard current
transmission output, with a load capacity of 750Ω (maximum), to facilitate
the acquisition of display instruments or DCS/PLC, to achieve long-distance
transmission of signals
Relay alarm output:
up to 6 channels support relay alarm output,
contact capacity 1A@250VAC/1A@30VDC (maximum contact capacity can
be customized), can be configured on the upper limit, upper limit, lower
limit, lower limit alarm
Communication printing
Communication interface:
RS232C and RS485 two kinds of
communication interface for users to choose, support simultaneous use of
RS232C and RS485, support simultaneous use of RS232C and RS232C,
support Modbus RTU protocol, baud rate
----
(1200, 4800, 9600, 19200,
38400, 57600)
Print interface:
RS232C directly connected to a micro printer, baud rate
1200
Recording function
Recording capacity:
64/128/192/248MB (FLASH capacity is optional).
Recording interval:
1 second to 240 seconds, in 11 steps.
1/2/4/8/12/24/36/60/120/180/240 seconds selectable.
Recording time:
the length of the recording time is related to the FLASH
memory capacity, the number of input points, and the recording interval,
and the calculation formula is as follows (the units substituted into the
value should be consistent with the formula)
Recorded Days
FlASH Capacity (MB)
1024
1024
Recording Interval (second)
Day
Number of channels
16
24
3600
Summary of Contents for HT R71