- 42 -
④
Recording interval:
can be set as 1/2/4/8/12/24/36/60/120/180/240
seconds. The larger the recording interval, the longer the recording time,
and vice versa, the smaller the recording interval, the shorter the recording
time. In general, when the measured signal changes quickly, the recording
interval should be chosen smaller. On the contrary, when the measured
signal changes more slowly, the recording interval can be chosen to be
larger.
⑤
Key sound:
The system allows/prohibits the buzzer to make sound
when key operation.
⑥
Automatic switching:
Cycle the time period value of each display
combination screen.
⑦
System maintenance:
[Clear power-down list] or [Clear alarm list] or
[Restore default settings] and other system maintenance functions, press
[Confirmation key] to enter the 'System maintenance' parameter selection.
Once the system maintenance operation is confirmed, the relevant data will
be cleared or the factory settings will be restored, the process is irreversible,
please operate carefully.
⑧
PID adjustment:
Allow/prohibit the adjustable function of PID screen
parameters, i.e. when it is turned on, the parameters in the PID screen are
allowed to be adjusted, and vice versa is not allowed.
⑨
Shift report setting:
Set the shift and starting time of the accumulated
shift report, starting from the setting time of shift 1, in the order of shift
1→shift 2→shift 3→shift 1, and ending when the next shift 1 starts, for a
cycle of 24 hours in total. The shift report setting should follow the
following principles: Shift 1 < Shift 2 < Shift 3, and the accumulation of each
channel needs to be cleared after setting the shift report.
⑩
Operation:
Press [F5] to quickly exit the system configuration screen.
Press [Left key] or [Right key] to move the cursor.
Press [Up key] or [Down key] to switch the selection or
adjust the value.
Summary of Contents for HT R71
Page 22: ... 22 4 2 Usage Patterns ...