Step 1:
Push the meter into the mounting hole from the front of the
mounting panel (please use a steel plate). Where the thickness of the
mounting panel is (1.5~4.0) mm.
Step 2:
Use the mounting bracket that comes with the meter to install it
as shown above (two brackets are installed on the left and right sides of the
meter, and the screws used for the mounting bracket of the instrument
panel are M3 standard screws).
Step 3:
After the meter body is installed, the signal and power cables can
be connected
。
Note
When installing this instrument, please do not tilt it from side to side, and
try to install it horizontally (it can be tilted back <30°).
3.4 Instrument Wiring
3.4.1 Wiring method
It is recommended to use U-type crimp terminals with insulating sleeve
(M3 screws for power and signal terminals).
To enhance the safety of the instrument, observe the following warnings
when wiring.
Note
To prevent electric shock, make sure the supply power is disconnected
before wiring.
To prevent fire, use double-insulated wire (wire with a cross-sectional area
≥ 0.75 mm² is recommended for the power supply wire; wire with a
cross-sectional area ≥ 0.5 mm² is required for the relay output wire with
high voltage resistance).
Summary of Contents for HT R71