- 52 -
5.6 Communication Configuration
Move the cursor to the [Communication] entry in the [Configuration]
screen and press [Confirmation Key] to enter [Communication
Configuration], the screen is introduced as shown in the figure.
①
Online mode:
It includes both PC and printer. To enter the print
configuration, you must set the online mode to printer first.
②
Online address:
The communication online address is used to
distinguish the instrument when it forms a network, it is the identification
of the instrument in the network. It is used by the host software to access
the instrument; the local address can be set between 001 and 255 in the
same communication network, and cannot be repeated.
③
Baud rate:
The baud rate can not be changed when the communication
mode is 'printer' mode, and the baud rate can be selected when 'PC' mode
(1200, 4800, 9600, 19200, 38400, 57600).
④
Calibration mode:
no parity / odd parity / even parity, default odd
parity.
⑤
Stop bit:
2 bits / 1 bit, default 2 bits.
⑥
Operation:
Press [F5] to quickly exit the communication configuration.