13
Unit weight Calling-out
1) Press
UWS/ITEM
key. The window displays as shown on the right
(Press
M-/HI
to clear the value in the unit price screen firstly).
.
2) Input the number of unit weight. Press key
e
/MR
or
TARE/
f
to shift and key
M-/HI
or
PRINT/LO
to modify the value (or via numeric keys)
3) Press key
M+/SAVE
to confirm called-out weight and return to the weighing mode. Press
key
ZERO/ESC
to exit and return to weighing mode.
7-4-2 Sample Counting and ACAI (when the ACAI function is on)
Sample Counting
1) Place samples onto the weighing pan (or into a tared container) and input the quantity
with numeric keys, then press key
SMPL
.
The unit weight is calculated, with
Symbol “ ”
pointing at “ACAI”
.
Note:
●
The larger the sample size, the more accurate unit weight.
●
Symbol “ ” points at “SMPL” when calculated unit weight is lower than 4/5 of scale
division or the quantity of sample is less than 10 pieces.
2) Remove the samples and put the load on, the scale begins to count.
3) Press key
M-/HI
to exit and return to the weighing mode.
ACAI
Automatic Counting Accuracy Improvement (ACAI) results in a more accurate count by
increasing the reference weight without the need to count additional parts. A higher reference
weight is important when there is a risk of inconsistent piece weights or if the reference weight
is close to the minimum. ACAI uses an initial averaging unit weight to count additional pieces
that are placed on the scale. After a few seconds, the scale gives a beep as the new higher
reference weight is used to recalculate the averaging unit weight. The process can be
repeated as long as the additional weight is less than the previous reference weight. Once
this limit is exceeded, ACAI is turned off.
Manual ACAI closing: ACAI acts automatically after sampling when ACAI indication appears.
Long press
SMPL
key to turn off ACAI when ACAI indication disappears. ACAI re-acts when