8
10)
SET09
printing modes setting:
key
,
stable
,
contin
contin
:
Printing continually
stable
:
Printing automatically when get a stable weight (more than 20d)
key
:
Printing by pressing keys
11)
SET10
printing format setting: prt01~prt03
Use the numeric key to set the format directly.
12)
SET11
filtering setting:
1
,
2
,
3
,
4
Set the filtering level in which the stable indication turns on .The higher the setting, the
slower stabilization time.
13)
SET12
Precondition for zeroing or taring setting:
stable, auto, always
stable
:
Taring or zeroing can be not action without stable indication after pressing the key
TARE
or
ZERO
always
:
Taring or zeroing can be action without stable indication after pressing the key
TARE
or
ZERO
auto
:
Press key
TARE
or
ZERO
, but it works only when get the stable indication.
14)
SET13
weight memory parameter:
on
,
off
On:
display the last weight when powering on again
Off:
do not display the last weight when powering on again
15)
SET14
the kinds of checking:
Qty.ck
,
Wt.ck
Qty.ck:
start amount checking
Wt.ck:
start weighing checking
16)
SET15
ACAI function setting: ON, OFF
17)
SET16
parameter initialization
press key
G N/SET
then key
M+/SAVE
to start initialization. It is finished when
displaying OK.
6. Single and Multi-point Calibration
Note:
Before calibration, please set the capacity first. The unit used in calibration is the one
that has been set before. During the calibration procedure, LONG press
ZERO/ESC
to
return to normal weighing mode without saving.
Here we take 3kg/0.2g as an example