12
and return to normal weighing mode , press key
ZERO/ESC
.
7-6 Printer initialization by the indicator operation
1. Press
M-/SET
while powering on to enter parameter setting. Use key
/MR
or
TARE/
to shift to P09
2. Press
M-/SET
to enter the external device setting. Press
/MR
or
TARE/
to choose
the printer model
3. After choosing one printer model, press
M-/SET
and the window will show “UNSUP” or
“INIT?”
z
“UNSUP” means the printer can’t be initialized by scale. Press
ZERO/ESC
to return.
z
“init?” means the printer is should be initialized. Press
MR/SAVE
to initialize the printer.
When the initialization is finished, the window will show “ok”. And then displays the
printer model. Press
ZERO/ESC
to return.
7-7 Input commands
Connect the scale to computer. Run serial port debugging program on the computer. Input the
capital number “Z”, “T”, “R” ,“C”,“P” in the sending area, and the indicator can conduct the
corresponding actions.
Z=zero
T=tare
C=cancel tare
R/P=reading / print
8. Serial Interface
If external interface is needed, please select the proper RS 232 board or Relay port firstly, and
only when this board is adopted, the functions can be enabled.
Note: RTC is unavailable.
8-1 RS-232 Diagram