Step 6: Check the injection head
•
Check if the injection head is sitting correctly and if the safety cylinder
around the injection head (marked yellow) is moving smoothly.
Step 7: Flush
•
Fit a bottle with rinsing/cleaning fluid into the holder and rinse until a clean and continuous
jet is coming out of the injection head. (Use rinsing program, see 10.1)
Step 8: Check functionality
•
If the injection head is not operating properly (no clean and continuous jet of rinsing fluid is
coming out), please repeat cylinder service.
Step 9: Reset Service Counter
•
After successful completion of cylinder service, please
reset service counter.
•
Starting from the initial screen, push the right button three
times until reaching menu point “Service Counter”.
•
Press button “Reset” for at least 3 seconds.
•
The counter returns to "0".
11.2.
Maintenance and repair service
For certain problem reports, maintenance or repair in an approved service point is necessary. The
message “Factory Service” will appear on the EPIG® display. The EPIG® device has to be sent to an
approved service point. Details can be found in the table on the following page. Please contact Henke-
Sass, Wolf to arrange return of the device and fill out the supplied return form.
CAUTION:
In case of a necessary return of the device to the manufacturer, please
remove
the batteries from the case
(as per dangerous goods regulation of all common forwarders)!
To learn more on what sort of failure occurred, download the EPIG® App for iOS or Android and run a
self-diagnosis by starting the App, connecting the App to your device and thereafter pushing the button
“Run diagnosis” in the App.
EN - EPIG User manual
17
Summary of Contents for EPIG
Page 24: ...23 17 EC declaration EN EPIG User manual 24 ...
Page 48: ...24 17 EC verklaring NL EPIG Handleiding 48 ...
Page 72: ...24 17 EC Erklärung DE EPIG Bedienungsanleitung 72 ...
Page 95: ...23 17 Déclaration CE FR EPIG Manuel d utilisation 95 ...
Page 119: ...24 17 Declaración CE ES EPIG Manual de usuario 119 ...