Troubleshooting
Operation:
Problem
Cause
Solution
Injection is not possible.
Contaminated injection head.
Clean injection head (see 10.2).
Malfunctioning cylinder.
Perform Cylinder Service (see 11.1).
Application error.
Place injection head perpendicularly
on the piglet skin (see 9).
When cleaning internally, after 10 cycles, no
rinsing fluid comes out of the injection
heads.
Pump is not working.
Return the EPIG® device to supplier.
Vial not properly positioned inside the
bottle holder.
Remove the vial and re-position it on
the spike inside the bottle holder.
Contaminated injection head.
Clean injection head (see 10.2).
The safety sleeve around the injection head
remains jammed following vaccination.
Contaminated injection head.
Clean injection head (see 10.2).
Possible damage to the injection head or to
the safety cylinder.
Release the trigger immediately,
remove the battery and perform
exchange of injection head using the
appropriate spare part (not
included).
A lot of vaccine on the skin of the animal and
no puncture visible (as mentioned in chapter
9,
it is normal that a small residue of
vaccine will remain on the piglet’s skin).
Air inside the tubing system.
Rinse the device with rinsing fluid.
During the pump process, hold the
device horizontally, during the
rinsing process, hold the device
vertically (pointing upwards). Repeat
this procedure 10 times.
Display Indication:
Problem
Cause
Solution
"Cylinder Service"
The service counter records 12,000 shots: a
service is required.
Perform Cylinder Service. Please see
Chapter 11.1.
“Low Battery Please Charge”
The battery has to be charged. Vaccination
is still possible.
Fit a charged battery into the EPIG®
device and recharge the empty one.
“Load Process”
Weak battery.
Insert a fully charged battery.
Cylinder defective.
Perform “Cylinder Service” (see
Chapter 11.1).
Repair is needed.
Contact supplier.
“Air detected”
Air was drawn into the device.
Rinse device three (3) times.
“Please release nozzle”
Contact with piglet after injection was kept
upright.
Release device from the piglet.
“Warning! Early Release”
Device was released from the piglet too
early.
Repeat the injection properly.
CAUTION:
In case of a necessary return of the device to the manufacturer, please
remove
the batteries from the case
(as per dangerous goods regulation of all common forwarders)!
EN - EPIG User manual
22
Summary of Contents for EPIG
Page 24: ...23 17 EC declaration EN EPIG User manual 24 ...
Page 48: ...24 17 EC verklaring NL EPIG Handleiding 48 ...
Page 72: ...24 17 EC Erklärung DE EPIG Bedienungsanleitung 72 ...
Page 95: ...23 17 Déclaration CE FR EPIG Manuel d utilisation 95 ...
Page 119: ...24 17 Declaración CE ES EPIG Manual de usuario 119 ...