To disconnect a bottle from the device, proceed in the reverse order. Always make sure to have the
draw-off spike of the device pointing downwards when connecting or disconnecting a bottle.
Otherwise, vaccine will be spilled over the device while a bottle is being connected to or disconnected
from the bottle mount.
If using the tubing collar to connect a larger bottle via tubing,
prefill the tubing
(!) prior to attaching it
to the tubing collar. When the tube starts to fill, kink the tube just before the end to prevent loss of
vaccine. Then quickly connect the tubing to the adapter.
6.
Device Start
Once the vaccine bottle is connected, insert the battery and the device will turn on, showing both the
Henke-Sass Wolf logo and the EPIG® logo on the display. Both logos are shown for 1 second.
If the battery is sufficiently charged:
The display will switch initially to the User Menu, unless the
device is in a loaded condition (cylinder filled with vaccine / rinsing fluid) – in which case the display
will show “! Loaded Device ! Press any button” and the LED bars turn red.
By pressing one of the display buttons, the device will rinse once, to unload the device. The liquid will
be sprayed out through the injection head.
Afterwards, the device will switch to the initial display of the User Menu (see chapter 7).
Before pushing the trigger for the first time, make your settings in the User Menu.
EN - EPIG User manual
9
Summary of Contents for EPIG
Page 24: ...23 17 EC declaration EN EPIG User manual 24 ...
Page 48: ...24 17 EC verklaring NL EPIG Handleiding 48 ...
Page 72: ...24 17 EC Erklärung DE EPIG Bedienungsanleitung 72 ...
Page 95: ...23 17 Déclaration CE FR EPIG Manuel d utilisation 95 ...
Page 119: ...24 17 Declaración CE ES EPIG Manual de usuario 119 ...