1.
Terminate the programme.
Press the Press the STOP button.
The programme will be terminated and the vacuum chamber is decompressed.
6.1.6 Changing the Programme Settings
This section describes the units and limits of the parameters and how parameters can be adjusted.
See Operating Elements on page 19 for an overview of the operating elements of the 1-PCS.
1.
Use the Cursor key to scroll to the desired parameter.
The LED in front of the selected function will light up.
2.
Press the – / STOP button and the + / VACUUM STOP button to adjust the value.
The – / STOP button and the + / VACUUM STOP button must be pressed and held for several
seconds to adjust the value. This prevents the accidental changing of settings.
3.
Press the Cursor key to activate the new parameter.
6.1.6.1 Vacuum
During the cycle, the air is removed from the chamber until the set time or percentage has been
reached.
1.
Use the Cursor key to scroll to the parameter Vacuum.
The LED in front of the selected function will light up.
2.
Press the + / VACUUM STOP button to adjust the value.
The – / STOP button and the + / VACUUM STOP button must be pressed and held for several
seconds to adjust the value. This prevents the accidental changing of settings.
3.
Press the Cursor key to activate the new parameter.
6.1.6.2 time (only for sensor control)
When air inclusions appear in the product, it may be desirable to prolong the vacuum time after
reaching the maximum vacuum. This process is meant to let enclosed air escape from the product.
The time is set in seconds. If a time is set, a dot will appear in the lower right
corner of the parameter display.
1.
Use the Cursor key to scroll to the parameter Vacuum.
The LED in front of the selected function will light up.
2.
Press the + / VACUUM STOP button to adjust the value to the maximum vacuum percentage
of 99%.
The + / VACUUM STOP button must be pressed and held for several seconds to adjust the
value. This prevents the accidental changing of settings.
3.
Press the Cursor key once to select the time.
The display shows O.
The LED stays lit in front of VACUUM.
4.
Press the buttons – / STOP and + / VACUUM STOP to change the value in seconds.
A dot will appear in the lower right corner of the parameter display when the value is set.
5.
Wait a few seconds or press the Cursor key to activate the new parameter.
Operation 21