background image

 

Hettich AG, Seestrasse 204a, 8806 Baech, Switzerland 

Operator Manual_DancerPlus_EN_rev, 1.1 

 

 

 

Page 19 

2.3

 

Installation 

2.3.1

 

Location 

The DancerPlus must be structurally stable for operation.  
In order to ensure a sufficient fresh air supply, the ventilation slots (Fig. 6 no. 1 or Fig. 8 no. 3) 
must not be blocked or obstructed with materials, paper, a folder etc., and a minimum distance 
of 20 cm from an impediment must be maintained. The fan should always work in parallel with 
the IR lamp (Fig. 7 no. 2) when the lid is closed.  
 
The DancerPlus and the vacuum pump shake when in operation, which means that these devices 
must not be on the same working surface as, for example, a microbalance, microscope, HPLC, 
sensitive analysis devices, sample and cell culture storage areas, etc.  

 

2.3.2

 

Operation with inert gas, vacuum pump, cold trap 

For operation with inert gas, e.g. nitrogen, the inert gas line is connected at the release inlet (Fig. 6 no. 3) in order 
to  ventilate  or  fill  up  the  vacuum  chamber  (Fig.  7)  during  the  vacuum  relief  process  with  inert  gas.  This  is 
recommended during the evaporation of solvents, e.g. ether, benzene etc. with a low flashpoint or ignition point. 

 
Install a pressure reduction valve with a manometer featuring a display range of 0 bar to 6 bar 
and set the pressure to 2 bar (29psi). 
As nitrogen is heavier than  air, open the cover carefully and slowly so that the samples in the 
vacuum chamber remain shielded in the inert gas build-up. 
The  security  guidelines  relating  to  inert  gas  in  general,  and  in  particular  following  a  power 
interruption or defect, must also be observed. 

 

Diaphragm and rotary vane pumps are all suitable for operation with the DancerPlus. We recommend that you 
use a Vacuubrand® PC3003 Vario diaphragm pump for the careful constriction or evaporation of solvents. This 
will  allow  pumping  to  be  carried  out  at  time-controlled  gradients,  as  well  as  with  automatic  boiling  pressure 
detection and tracking. A rotary vane pump is better suited for the evaporation of solvents such as DMSO, or for 
the drying of powders or silica gel. 
 

A  diaphragm  pump  and  rotary  vane  pump  with  a  switchover  valve  may  be  used  for  the 
evaporation of various solvents and for different working processes. 

 
We  recommend pre-connecting a cold trap, e.g. the  PoleStar, to aid the vacuum pump  in the  evaporation of 
solvents. This not only saves time during the evaporation process, but also provides relief to the vacuum pump, 
ambient  air,  the  persons  present  in  the  area  and  the  environment  due  to  the  hazardous  substances  being 
effectively contained. 
 
The  vacuum  pump  should  be  installed  opposite  the  PoleStar  cold  trap  at  a  slight  increase  in  height,  and  is 
connected to the tube connection. The operating instructions of the PoleStar cold trap and the vacuum pump 
must be observed. 
 

2.3.3

 

When using a Vacuubrand® diaphragm pump and PoleStar cold trap: 

Fit the external vacuum sensor (Fig. 9) with the corresponding fittings so that the sensor protrudes 
from above with an angle of 45° ±15°; this will prevent condensation and liquid build-ups on the 
sensor. This will result in an incorrect pressure measurement. 
Fit the metal tube of the DancerPlus vacuum connection to prevent liquid build-ups or siphons in 
the metal tube 

 

 

Summary of Contents for DancerPlus

Page 1: ...Hettich AG Seestrasse 204a 8806 Baech Switzerland Operator Manual_DancerPlus_EN_rev 1 1 Page 1 Operating manual Infrared vortex evaporator DancerPlus...

Page 2: ...e 2 Manufactured by Hettich AG Seestrasse 204a CH 8806 Baech Switzerland phone 41 0 44 786 80 20 info hettich ch www hettich ch 2020 by Hettich AG All rights reserved No part of this document may be r...

Page 3: ...6 Jumo temperature controller 16 2 2 7 Label rating plate 16 2 2 8 Sample racks 17 2 3 Installation 19 2 3 1 Location 19 2 3 2 Operation with inert gas vacuum pump cold trap 19 2 3 3 When using a Vacu...

Page 4: ...ng hour display 23 3 0 6 3 Low high temperature alarm 24 3 0 6 4 Temperature value offset value 24 3 0 6 5 Temperature offset value 24 3 1 Cleaning carried out by the user 25 3 1 1 Precautions 25 3 1...

Page 5: ...nd reinstalled in a different location these operating instructions must be included Additional documents which you must observe the separate operating instructions for the vacuum pump the separate op...

Page 6: ...l 1 4301 10088 2 304 V2A shaking plate 1 4301 10088 2 304 V2A vacuum port DN 25 ventilation port 6mm hose connection max 3bar 43psi place of installation ambient conditions EN IEC61010 1 altitude ambi...

Page 7: ...related information and identifies potentially dangerous situations Non observance of this information may lead to property damage or personal injury Symbol in this document This symbol identifies imp...

Page 8: ...the DancerPlus to a residual current circuit breaker with a shockproof socket only use a power cable that fulfils the regulations and is in flawless condition Defective or inadequate earthing may have...

Page 9: ...device alone or in the system and have familiarised themselves with these operating instructions must be aware of its applicable regulations guidelines and health and safety specifications and instruc...

Page 10: ...salt may cause pitting corrosion and crevice corrosion hydrofluoric acid HF mine water urine iodine J2 iodine tincture potassium bifluoride KHF2 potassium bromide KBr potassium chloride KCl potassium...

Page 11: ...ted separately from and outside the DancerPlus unless the rack transportation lock is used If the device has to be transported away from its location e g to be relocated or for repairs at the manufact...

Page 12: ...s 2 2 1 Front view 1 2 3 10 Fig 5 4 5 6 7 8 9 11 12 Legend 1 Cover 6 IR lamp switch 2 Viewing window 7 Temperature controller 3 Cover handle 8 Shaker switch 4 Release valve switch 9 Shaker speed adjus...

Page 13: ...cerPlus_EN_rev 1 1 Page 13 2 2 2 Rear view 1 2 7 Fig 6 8 3 4 5 6 Legend 1 Ventilation slots 5 Label Rating plate 2 Lid connection cable IR lamp 6 Mains inlet device socket 3 Release inlet connection r...

Page 14: ...cerPlus_EN_rev 1 1 Page 14 2 2 3 View with open lid Fig 7 1 2 3 5 4 8 6 7 7 Legend 1 Borosilicate glass panel 5 Vacuum connection outlet 2 IR lamps and reflector 6 Temperature sensor 3 Gas spring 7 Re...

Page 15: ...egend 1 Borosilicate glass panel securing screws depicted on the left hand side the right hand side is identical 2 Folding handle 3 Ventilation slots 2 2 5 Vacuum sensor fitting Fig 9 1 2 Legend 1 Ext...

Page 16: ...ncerPlus 2 Programming button temperature setpoint entry 3 LED illuminates until the target value has been reached 4 Increase value button 5 Decrease value button 6 Display 2 2 7 Label rating plate Fi...

Page 17: ...teed if original racks are used Self built racks or setups that are not approved by the manufacturer or its national representative will result in the voiding of the guarantee on the shaking plate and...

Page 18: ...ht or CD Z912 for 60mm height The R0121 rack for the DWP 3 pieces or microtiter plates are not suitable for the FAST option for the shaking plate When the rack is removed the temperature sensor must f...

Page 19: ...articular following a power interruption or defect must also be observed Diaphragm and rotary vane pumps are all suitable for operation with the DancerPlus We recommend that you use a Vacuubrand PC300...

Page 20: ...t explosion proof this means a special exhaust system is required 2 3 7 Installation of the rack and shaking plate Without the FAST option the rack is screwed to the shaking plate this ensures that th...

Page 21: ...is not a standard cable and thus may only be replaced with an equivalent cable 2 4 Conduct in the event of faults and irregularities The device may only be operated in flawless condition If as a user...

Page 22: ...is reduced or completely prevented The compartment heater is controlled in parallel to the IR lamp heating At temperatures of 70 C and above the safety device may switching hysteresis 3 0 C 5 0 C in c...

Page 23: ...To change these press the corresponding arrow button Fig 10 no 4 or no 5 Briefly pressing the arrow button will change the value in 0 1 C increments Continuously pressing the arrow button will change...

Page 24: ...hen press the P button three times The factory setting of the value is 90 0 C If the actual temperature exceeds the high temperature value the optical alarm display AL H and the current actual tempera...

Page 25: ...wever if highly aggressive liquids as described in chapter 1 8 are evaporated then we recommend daily cleaning For any stubborn contamination we recommend that you first soak the contaminated area wit...

Page 26: ...required Remove and clean the O ring Fig 7 no 4 with a glass cleaning agent or ethanol 70 and check the O ring for bends and cuts it may need to be replaced Also clean the O ring guide slot on the va...

Page 27: ...n using the FAST option Clean the O rings and the rack pins Fig 2 no 4 and no 5 check the O rings for brittleness and cracks and replace them if required After cleaning lubricate the O rings with the...

Page 28: ...lus is subject to the 2002 96 EC Waste Electrical and Electronic Equipment Directive WEEE and may not be disposed of with ordinary household or industrial waste The disposal guidelines of the individu...

Page 29: ...06 FAST rack pins for the shaking plate CD Z908 FAST feet mounting set for DWP racks 30mm height CD Z909 FAST handle set for block racks CD Z928 centering ring DN25 ISO KF CD Z929 centering ring DN16...

Page 30: ...N_rev 1 1 Page 30 4 0 Revision history Rev replaced Version Revision description Datum 1 1 1 0 CI and orthography updated several typing errors corrected several sections restructured chap 1 4 tech da...

Reviews: