background image

 

Hettich AG, Seestrasse 204a, 8806 Baech, Switzerland 

Operator Manual_DancerPlus_EN_rev, 1.1 

 

 

 

Page 21 

2.3.8

 

Mains inlet device socket and mains switch

 

Please check the mains connection values on the label (rating plate) (see Fig. 11 no.3) on the rear 
side of the DancerPlus (see Fig. 6 no. 5) with regard to the power outlet to be used and observe 
the  country-specific  regulations  (e.g.  in  Germany,  DIN  VDE  0100,  power  outlet  with  residual 
current device). In case of doubt, please contact your technical service department, an electrician 
or electricity plant  
 
The mains inlet device socket (Fig. 6 no. 7) is located on the rear side of the DancerPlus, and the 
supplied power supply cable is connected here. Actuate mains switch on (Fig. 6 no. 8) so that 

the  “O”  position  protrudes  and  the  DancerPlus  is  switched  on.  Switch  on  the  other  system 

devices such as the vacuum pump and, if required, the cold trap. 
 

2.3.9

 

Lid connection cable 

Ensure that the lid connection cable (Fig. 6 no. 2) is not damaged and cannot be damaged, e.g. 
transportation damage, torn-out plastic cable gland, cable clamped between the basic device 
and the lid. 
Attention: this is not a standard cable and thus may only be replaced with an equivalent cable 
 

2.4

 

Conduct in the event of faults and irregularities 

The device may only be operated in flawless condition. If, as a user, you notice any irregularities, 
faults or damage, shut down the device immediately and inform your supervisor. 

 

2.5

 

Emergency switch-off of the DancerPlus 

Change the position of the power switch on the rear side of the device (Fig. 6 no. 8) and disconnect the power 
cable from the mains. 
 

If  the  DancerPlus  is  switched  off  or  disconnected  from  the  mains,  e.g.  following  a  power 
interruption, the release valve (release inlet) (Fig. 6 no. 3 and Fig. 5 no. 4) will automatically open, 
and if the vacuum pump is stopped or switched off, the lid may be opened after approx. 3 minutes 
in order to remove the samples from the DancerPlus.  
 
Depending  on  the  type  of  operation,  the  vacuum  chamber  surfaces  (Fig.  7),  the  temperature 
sensor, the rack and the sample vials / tubes contained within may still be hot after switching off 
or a power interruption. They may cause burns if touched. Use temperature-resistant protective 
gloves or allow the DancerPlus to initially cool down after switching off.  
Open the lid to allow it to cool down quicker. 
 
When dealing with samples, take account any potential biohazards and wear protective gloves 
and goggles. 
 

 

 

Summary of Contents for DancerPlus

Page 1: ...Hettich AG Seestrasse 204a 8806 Baech Switzerland Operator Manual_DancerPlus_EN_rev 1 1 Page 1 Operating manual Infrared vortex evaporator DancerPlus...

Page 2: ...e 2 Manufactured by Hettich AG Seestrasse 204a CH 8806 Baech Switzerland phone 41 0 44 786 80 20 info hettich ch www hettich ch 2020 by Hettich AG All rights reserved No part of this document may be r...

Page 3: ...6 Jumo temperature controller 16 2 2 7 Label rating plate 16 2 2 8 Sample racks 17 2 3 Installation 19 2 3 1 Location 19 2 3 2 Operation with inert gas vacuum pump cold trap 19 2 3 3 When using a Vacu...

Page 4: ...ng hour display 23 3 0 6 3 Low high temperature alarm 24 3 0 6 4 Temperature value offset value 24 3 0 6 5 Temperature offset value 24 3 1 Cleaning carried out by the user 25 3 1 1 Precautions 25 3 1...

Page 5: ...nd reinstalled in a different location these operating instructions must be included Additional documents which you must observe the separate operating instructions for the vacuum pump the separate op...

Page 6: ...l 1 4301 10088 2 304 V2A shaking plate 1 4301 10088 2 304 V2A vacuum port DN 25 ventilation port 6mm hose connection max 3bar 43psi place of installation ambient conditions EN IEC61010 1 altitude ambi...

Page 7: ...related information and identifies potentially dangerous situations Non observance of this information may lead to property damage or personal injury Symbol in this document This symbol identifies imp...

Page 8: ...the DancerPlus to a residual current circuit breaker with a shockproof socket only use a power cable that fulfils the regulations and is in flawless condition Defective or inadequate earthing may have...

Page 9: ...device alone or in the system and have familiarised themselves with these operating instructions must be aware of its applicable regulations guidelines and health and safety specifications and instruc...

Page 10: ...salt may cause pitting corrosion and crevice corrosion hydrofluoric acid HF mine water urine iodine J2 iodine tincture potassium bifluoride KHF2 potassium bromide KBr potassium chloride KCl potassium...

Page 11: ...ted separately from and outside the DancerPlus unless the rack transportation lock is used If the device has to be transported away from its location e g to be relocated or for repairs at the manufact...

Page 12: ...s 2 2 1 Front view 1 2 3 10 Fig 5 4 5 6 7 8 9 11 12 Legend 1 Cover 6 IR lamp switch 2 Viewing window 7 Temperature controller 3 Cover handle 8 Shaker switch 4 Release valve switch 9 Shaker speed adjus...

Page 13: ...cerPlus_EN_rev 1 1 Page 13 2 2 2 Rear view 1 2 7 Fig 6 8 3 4 5 6 Legend 1 Ventilation slots 5 Label Rating plate 2 Lid connection cable IR lamp 6 Mains inlet device socket 3 Release inlet connection r...

Page 14: ...cerPlus_EN_rev 1 1 Page 14 2 2 3 View with open lid Fig 7 1 2 3 5 4 8 6 7 7 Legend 1 Borosilicate glass panel 5 Vacuum connection outlet 2 IR lamps and reflector 6 Temperature sensor 3 Gas spring 7 Re...

Page 15: ...egend 1 Borosilicate glass panel securing screws depicted on the left hand side the right hand side is identical 2 Folding handle 3 Ventilation slots 2 2 5 Vacuum sensor fitting Fig 9 1 2 Legend 1 Ext...

Page 16: ...ncerPlus 2 Programming button temperature setpoint entry 3 LED illuminates until the target value has been reached 4 Increase value button 5 Decrease value button 6 Display 2 2 7 Label rating plate Fi...

Page 17: ...teed if original racks are used Self built racks or setups that are not approved by the manufacturer or its national representative will result in the voiding of the guarantee on the shaking plate and...

Page 18: ...ht or CD Z912 for 60mm height The R0121 rack for the DWP 3 pieces or microtiter plates are not suitable for the FAST option for the shaking plate When the rack is removed the temperature sensor must f...

Page 19: ...articular following a power interruption or defect must also be observed Diaphragm and rotary vane pumps are all suitable for operation with the DancerPlus We recommend that you use a Vacuubrand PC300...

Page 20: ...t explosion proof this means a special exhaust system is required 2 3 7 Installation of the rack and shaking plate Without the FAST option the rack is screwed to the shaking plate this ensures that th...

Page 21: ...is not a standard cable and thus may only be replaced with an equivalent cable 2 4 Conduct in the event of faults and irregularities The device may only be operated in flawless condition If as a user...

Page 22: ...is reduced or completely prevented The compartment heater is controlled in parallel to the IR lamp heating At temperatures of 70 C and above the safety device may switching hysteresis 3 0 C 5 0 C in c...

Page 23: ...To change these press the corresponding arrow button Fig 10 no 4 or no 5 Briefly pressing the arrow button will change the value in 0 1 C increments Continuously pressing the arrow button will change...

Page 24: ...hen press the P button three times The factory setting of the value is 90 0 C If the actual temperature exceeds the high temperature value the optical alarm display AL H and the current actual tempera...

Page 25: ...wever if highly aggressive liquids as described in chapter 1 8 are evaporated then we recommend daily cleaning For any stubborn contamination we recommend that you first soak the contaminated area wit...

Page 26: ...required Remove and clean the O ring Fig 7 no 4 with a glass cleaning agent or ethanol 70 and check the O ring for bends and cuts it may need to be replaced Also clean the O ring guide slot on the va...

Page 27: ...n using the FAST option Clean the O rings and the rack pins Fig 2 no 4 and no 5 check the O rings for brittleness and cracks and replace them if required After cleaning lubricate the O rings with the...

Page 28: ...lus is subject to the 2002 96 EC Waste Electrical and Electronic Equipment Directive WEEE and may not be disposed of with ordinary household or industrial waste The disposal guidelines of the individu...

Page 29: ...06 FAST rack pins for the shaking plate CD Z908 FAST feet mounting set for DWP racks 30mm height CD Z909 FAST handle set for block racks CD Z928 centering ring DN25 ISO KF CD Z929 centering ring DN16...

Page 30: ...N_rev 1 1 Page 30 4 0 Revision history Rev replaced Version Revision description Datum 1 1 1 0 CI and orthography updated several typing errors corrected several sections restructured chap 1 4 tech da...

Reviews: