Hettich AG, Seestrasse 204a, 8806 Baech, Switzerland
Operator Manual_DancerPlus_EN_rev, 1.1
Page 24
3.0.6.3
Low / high temperature alarm
To change the low
temperature alarm, press the “P” button three times and the display will alternate between the
value “AL.L” and the current low temperature alarm value. To change these, press the corresponding arrow button
and then press the “P” button four times. The fac
tory setting of the value is 9.0°C.
If the actual temperature falls below the low temperature value, the optical alarm display “AL.L” and the current
actual temperature value will appear alternately in the display. The non-changeable alarm hysteresis is 2.0°K.
The current actual temperature value must be 2.0°K higher than the set low temperature alarm
value so that the alarm will be deleted, and the display will no longer flash.
To change the high temperature alarm, press the “P” button four times and t
he display will alternate between the
value “AL.H” and the current low temperature alarm value. To change these, press the corresponding arrow
button and then press the “P” button three times. The factory setting of the value is 90.0°C.
If the actual temp
erature exceeds the high temperature value, the optical alarm display “AL.H” and the current
actual temperature value will appear alternately in the display. The non-changeable alarm hysteresis is 2.0°K.
The current actual temperature value must be 2.0°K lower than the set high temperature alarm
value so that the alarm will be deleted, and the display will no longer flash.
The alarm, whether low temperature or high temperature, will only appear optically.
The DancerPlus requires the option, DP.Z940, for an acoustic alarm.
3.0.6.4
Temperature value offset value
To change the alarm delay, press the “P” button five times and the display will alternate between the value “AL.d”
and the current alarm delay value. To change these, press the corresponding arrow button and then press the
“P” button two times. The factory setting of the value is 0 min.
3.0.6.5
Temperature offset value
Press the “P” button six times to read the temperature offset value. The value “0F.t” will appear alternately in the
display and the
current temperature sensor offset value will be displayed in hours (“xxh”). This value is merely
informative and can only be changed by an authorised technician.