background image

 

Hettich AG, Seestrasse 204a, 8806 Baech, Switzerland 

Operator Manual_DancerPlus_EN_rev, 1.1 

 

 

 

Page 10 

1.10

 

Resistance of the vacuum chamber 

When the following chemicals are evaporated, there is a risk of erosive surface corrosion, pitting, crevice corrosion 
or stress corrosion cracking according to our chrome steel supplier Nirosta®. This may lead to vacuum leakage. 
 
Aluminium chloride (AlCl3 * 6H2O), ammonium bifluoride (NH4HF2), ammonium chloride (NH4Cl), aniline 
hydrochloride (C6H5NH2HCl), antimony trichloride (SbCl3) 
barium chloride (BaCl2 *2H2O), blood (in the presence of salt, pitting corrosion and crevice corrosion may 
occur, particularly with porcine blood), bromine (Br2), bromine water 
calcium chloride (CaCl2 * 6H2O), calcium hypochloride (Ca(OCl)2 * 4H2O), chlorine (Cl2), chlorinated lime 
((3CaCl(OCl) * Ca(OH)2) * 5H2O), chloric acid (HClO3), chlorosulfuric acid (HSO3Cl), chlorine water, hydrogen 
chloride gas (HCl) 
iron (III) chloride (FeCl3), iron gall ink (inks containing salt may cause pitting corrosion and crevice corrosion) 
hydrofluoric acid (HF) 
mine water 
urine 
iodine (J2), iodine tincture 
potassium bifluoride (KHF2), potassium bromide (KBr), potassium chloride (KCl), potassium hypochloride (KClO), 
aqua regia (HCl + HNO3), copper (II) chloride (CuCl2 * 2H2O) 
magnesium chloride (MgCl2 * 6H2O), monochloroacetic acid (CH2ClCOOH) 
sodium chloride (NaCl), sodium chlorite (NaClO2), sodium hypochlorite (NaClO) 
nickel chloride (NiCl2 * 6H2O) 
photographic fixing bath, brine solution, pink salt ((NH4)2(SnCl6)) 
mercury (II) chloride (HgCl2) 
hydrochloric acid (HCl), pickled cabbage brine, sulphuric acid (H2SO4), seawater, mustard, silver bromide 
(AgBr), silver chloride (AgCl) 
toluene sulfonic chloramide sodium (CH3C6H4SO2NClNa * 3H2O), trichloroacetic acid (CCl3COOH) 
water, acidic (mine water, dependent on chloride concentration), zinc chloride (ZnCl2), tin (IV) chloride (SnCl4), 
tin ammonium hexachloride 
 
 

When evaporating the aforementioned solvents using the DancerPlus, we recommend thoroughly 
cleaning the vacuum chamber and rinsing it daily. 
This includes the tube connection system for the vacuum pump and, if required, for the cold trap, 
as well as the cold trap chamber. 
 

 

 

Summary of Contents for DancerPlus

Page 1: ...Hettich AG Seestrasse 204a 8806 Baech Switzerland Operator Manual_DancerPlus_EN_rev 1 1 Page 1 Operating manual Infrared vortex evaporator DancerPlus...

Page 2: ...e 2 Manufactured by Hettich AG Seestrasse 204a CH 8806 Baech Switzerland phone 41 0 44 786 80 20 info hettich ch www hettich ch 2020 by Hettich AG All rights reserved No part of this document may be r...

Page 3: ...6 Jumo temperature controller 16 2 2 7 Label rating plate 16 2 2 8 Sample racks 17 2 3 Installation 19 2 3 1 Location 19 2 3 2 Operation with inert gas vacuum pump cold trap 19 2 3 3 When using a Vacu...

Page 4: ...ng hour display 23 3 0 6 3 Low high temperature alarm 24 3 0 6 4 Temperature value offset value 24 3 0 6 5 Temperature offset value 24 3 1 Cleaning carried out by the user 25 3 1 1 Precautions 25 3 1...

Page 5: ...nd reinstalled in a different location these operating instructions must be included Additional documents which you must observe the separate operating instructions for the vacuum pump the separate op...

Page 6: ...l 1 4301 10088 2 304 V2A shaking plate 1 4301 10088 2 304 V2A vacuum port DN 25 ventilation port 6mm hose connection max 3bar 43psi place of installation ambient conditions EN IEC61010 1 altitude ambi...

Page 7: ...related information and identifies potentially dangerous situations Non observance of this information may lead to property damage or personal injury Symbol in this document This symbol identifies imp...

Page 8: ...the DancerPlus to a residual current circuit breaker with a shockproof socket only use a power cable that fulfils the regulations and is in flawless condition Defective or inadequate earthing may have...

Page 9: ...device alone or in the system and have familiarised themselves with these operating instructions must be aware of its applicable regulations guidelines and health and safety specifications and instruc...

Page 10: ...salt may cause pitting corrosion and crevice corrosion hydrofluoric acid HF mine water urine iodine J2 iodine tincture potassium bifluoride KHF2 potassium bromide KBr potassium chloride KCl potassium...

Page 11: ...ted separately from and outside the DancerPlus unless the rack transportation lock is used If the device has to be transported away from its location e g to be relocated or for repairs at the manufact...

Page 12: ...s 2 2 1 Front view 1 2 3 10 Fig 5 4 5 6 7 8 9 11 12 Legend 1 Cover 6 IR lamp switch 2 Viewing window 7 Temperature controller 3 Cover handle 8 Shaker switch 4 Release valve switch 9 Shaker speed adjus...

Page 13: ...cerPlus_EN_rev 1 1 Page 13 2 2 2 Rear view 1 2 7 Fig 6 8 3 4 5 6 Legend 1 Ventilation slots 5 Label Rating plate 2 Lid connection cable IR lamp 6 Mains inlet device socket 3 Release inlet connection r...

Page 14: ...cerPlus_EN_rev 1 1 Page 14 2 2 3 View with open lid Fig 7 1 2 3 5 4 8 6 7 7 Legend 1 Borosilicate glass panel 5 Vacuum connection outlet 2 IR lamps and reflector 6 Temperature sensor 3 Gas spring 7 Re...

Page 15: ...egend 1 Borosilicate glass panel securing screws depicted on the left hand side the right hand side is identical 2 Folding handle 3 Ventilation slots 2 2 5 Vacuum sensor fitting Fig 9 1 2 Legend 1 Ext...

Page 16: ...ncerPlus 2 Programming button temperature setpoint entry 3 LED illuminates until the target value has been reached 4 Increase value button 5 Decrease value button 6 Display 2 2 7 Label rating plate Fi...

Page 17: ...teed if original racks are used Self built racks or setups that are not approved by the manufacturer or its national representative will result in the voiding of the guarantee on the shaking plate and...

Page 18: ...ht or CD Z912 for 60mm height The R0121 rack for the DWP 3 pieces or microtiter plates are not suitable for the FAST option for the shaking plate When the rack is removed the temperature sensor must f...

Page 19: ...articular following a power interruption or defect must also be observed Diaphragm and rotary vane pumps are all suitable for operation with the DancerPlus We recommend that you use a Vacuubrand PC300...

Page 20: ...t explosion proof this means a special exhaust system is required 2 3 7 Installation of the rack and shaking plate Without the FAST option the rack is screwed to the shaking plate this ensures that th...

Page 21: ...is not a standard cable and thus may only be replaced with an equivalent cable 2 4 Conduct in the event of faults and irregularities The device may only be operated in flawless condition If as a user...

Page 22: ...is reduced or completely prevented The compartment heater is controlled in parallel to the IR lamp heating At temperatures of 70 C and above the safety device may switching hysteresis 3 0 C 5 0 C in c...

Page 23: ...To change these press the corresponding arrow button Fig 10 no 4 or no 5 Briefly pressing the arrow button will change the value in 0 1 C increments Continuously pressing the arrow button will change...

Page 24: ...hen press the P button three times The factory setting of the value is 90 0 C If the actual temperature exceeds the high temperature value the optical alarm display AL H and the current actual tempera...

Page 25: ...wever if highly aggressive liquids as described in chapter 1 8 are evaporated then we recommend daily cleaning For any stubborn contamination we recommend that you first soak the contaminated area wit...

Page 26: ...required Remove and clean the O ring Fig 7 no 4 with a glass cleaning agent or ethanol 70 and check the O ring for bends and cuts it may need to be replaced Also clean the O ring guide slot on the va...

Page 27: ...n using the FAST option Clean the O rings and the rack pins Fig 2 no 4 and no 5 check the O rings for brittleness and cracks and replace them if required After cleaning lubricate the O rings with the...

Page 28: ...lus is subject to the 2002 96 EC Waste Electrical and Electronic Equipment Directive WEEE and may not be disposed of with ordinary household or industrial waste The disposal guidelines of the individu...

Page 29: ...06 FAST rack pins for the shaking plate CD Z908 FAST feet mounting set for DWP racks 30mm height CD Z909 FAST handle set for block racks CD Z928 centering ring DN25 ISO KF CD Z929 centering ring DN16...

Page 30: ...N_rev 1 1 Page 30 4 0 Revision history Rev replaced Version Revision description Datum 1 1 1 0 CI and orthography updated several typing errors corrected several sections restructured chap 1 4 tech da...