Tracker tools
H000002007-RDS User Manual
│
Version 5.0.0 (2018-09-17)
│96
E.3
SETTINGS
To access the Tracker settings, click on the button
Measurement profile
The measurement profile enables the user to select the adequate measurement behaviour according the
application needs.
Stationary
(default)
This is the standard manual way to measure: point by point, each point is measured
independently.
Continuous time
Frequency controlled scan probing: points are automatically triggered on a frequency
defined with a specified period.
The space between each point depends on the speed of movement: the faster it is,
the more spaced the points are.
The delay between each point can then be specified (default 500ms)
Continuous
distance
Interval controlled scan probing: points are automatically triggered at a specified
distance from the previous point.
The space between each point is constant and does not depend on the speed of
movement.
The distance between each point can then be specified (default 1mm)
Another probing option is available on the home screen to enable the triggering of the point
based on the stability of the probe: see next chapter.
Accuracy
The accuracy defines the time of taking points: the slower it is, the more accurate.
Recommended usage
Fast
With this mode, the delay for each point is
reduced, however, it means less accurate.
Typically
achieves
specified
accuracy in controlled environment.
Standard
(default)
This is the standard mode, a compromise
between speed and accuracy.
Achieves specified accuracy.
Precise
With this mode, the time to take each point is
longer, but it provides a higher accuracy.
For sensors checks or compensations.
Highest accuracy. Sensor checks
and compensations.