Operation
TP-R-400
26
2.2.4.3
Possible settings in the "Setup” menu
The following table shows the setting options for each parameter in the
"Setup”
menu.
Parameter
Possible Setting
Description
Mode
Probe
Using the system as a touch probe
Toolset
Using the system as a tool setter
Channel
*
)
1
Receiver channel 1
2
Receiver channel 2
Activation Code
A
⁞
H
Activation code of the touch probe
Safety
Fast
Low interference safety
Normal
Average interference safety (Standard)
Secu.2
High interference safety (Trigger Evaluation 2 ms)
Secu.10
High interference safety (Trigger Evaluation 10 ms)
Secu.20
High interference safety (Trigger Evaluation 20 ms)
Receiver Serial
XXXXX
Five-digit serial number of the receiver to be paired
*
)
Use of the receiver channel is only possible in conjunction with RWR95.51. For other receiver types, this setting
must always be set to
"1"
.
2.2.4.4
Setting Procedure
1. Press the button (Fig. 12) on the back of the touch probe for approx. 2 s to open the menu. Then immediately
press the button twice to confirm.
The display (Fig. 12) shows brief operating information for the touch probe menu and then opens the main
menu.
Entering the settings
2. Scroll to the
"Setup"
menu item by briefly pressing the button several times and select it by pressing the button
twice.
NOTICE
Risk of measuring deviations!
•
The touch probe needs to be urgently recalibrated to avoid measuring deviations after the interference safety
setting (parameter
"Safety"
) has been changed!
Button
Display
Summary of Contents for TP-R-400
Page 1: ......