TP-R-400
Operation
29
2.2.6
Adjusting the touch probe in the tool holder with a 90° adapter
1. Loosen adjustment screws "A" (4x) then re-tighten with light force.
2. Loosen clamping screws "C" (2x) then re-tighten with light force.
3. Use cap head screw AF4 to adjust the angular position of the touch probe (±5°).
4. Tighten the clamping screws "C" (2x).
5. Tighten the adjustment screws "A" (4x) against each other.
6. Calibrate the touch probe.
Adjusting the touch probe in the tool holder with a 90° adapter
Air jet with swivel screw
fitting (91.00-SWV-XX)
Hexagon key AF4 mm
(3489)
Angle adapter for tool holder
VDI (D20-VDI-90)
up
down
up
down
±5°