Total Station Manual
-
171 -
pressF4(OK).
○
5 Use fine tuning to set
the vertical angle as
V1+3’, focus on the
reticle center accurately,
wait for stable value,
press [F4](OK).
[F4]
[Adjust Tilt X]
OK
HA :
VA :
Tilt:
偏置模式 :
342°11′59″
F2 down 3′
269°29′46″
91
○
6 After finishing, it will
display the results, press
[F4](OK), save and back
to menu.
[F4]
[Adjust Tilt X]
OK
HA :
VA :
Tilt :
偏置模式 :
342°11′59″
Xk:33.0859 X0:-55
269°29′46″
100
1.4
Instrument constant setting
In Tools Menu, choose “1 Adjust”, and then press [F4 to enter “Const. Setting”.
Press [F4](OK) to save after editing the constants.
[Const. Setting]
OK
Add Const. :
Mul. Const.:
偏置模式 :
-9
0
1.5
Factory setting
In Tools Menu, choose “1 Adjust”, and then press [5] to enter “Factory Setting”.
If you need to reset the instrument parameters to factory state, you can use this
function, press key [F4] (Yes) and then the instrument will auto power off.