DTS-60A / DTS-100A O
PERATORS
M
ANUAL
STEP 13
STEP 14
5.0
kV
STIR
NOTE
While the test process is enabled the user must
not open the LID. If this happens...
...the system detect it and immediately cuts off
the output voltage.
The Lid open warning is displayed.
After closing the Lid, the current test is discarded
and the system will continue from STEP 3.
?
STAGE
1
HN on
TEST RESULTS
PRESS ANY KEY...
NUM. BD: 10
MEAN: 48.24kV
STD: 02.49kV
VOLTAGE:
0.5 kV/s
VOLTAGE:
0.5 kV/s
TEST:SIMPLE TEST
The Simple Test Results are shown.
The number of breakdowns, mean breakdown
voltage, and standard deviation are displayed
A Question Message is shown asking if a printout
is desired.
After the printout operation the system will return
to STEP 1 to allow another Simple Test or entry
into the Main Menu.
12:10 01/08/16 65°F
Print
Results?
TEST RESULTS
TEST:SIMPLE TEST
NUM. BD: 10
12:10 01/08/16 65°F
12:10 01/08/16 65°F
YES
TEST:SIMPLE TEST
TEST:SIMPLE TEST
TIMER:00:00:10
TIMER:00:00:10
Lid is Open!
Figure 3-8d –
Simple Test Example
35
Summary of Contents for DTS-100A