Step 6: Prepare the Master

The master is the head box and tank combo.

Place the master in its permanent location.

Ensure that the ball valve is in the open position. 

the tank MUST 

NOT be under pressure.

remove the hex nut cap and install a bushing in the opening. The 

bushing will be used to connect the braided steel line that runs 

between The master and Drone tanks.

9

NOTE

Don’t plug the system in until the 

last step. No power is needed yet!

side view

Remove hex nut “cap”

Summary of Contents for HVO GO+GO NAKED

Page 1: ...Setup Guide for your HVO 60 60 Naked Drone 1 ...

Page 2: ...n Regulator 4 Get the Tools You ll Need 5 Unpack the crate 6 Prepare the Master 7 Prepare the Naked Drone 8 Prepare the Oxygen Concentrators 9 Connect the components 10 Power Up You can do steps 1 4 while your system is in transit 2 Click this link to watch a video about the basic HVO setup ...

Page 3: ...oxygen and which have not been treated for oxygen use may be extremely hazardous For example propane tanks should never under any circumstances be used as oxygen storage tanks HVO tanks are designed to store oxygen generated by PSA oxygen concentrator systems only The pressure in these tanks should not exceed 150 psi HVO tanks are equipped with a 200 psi blow off valve Do not tamper with or replac...

Page 4: ... must be 45 90 If you plan to build shelving or cabinets for your HVO system Make sure the enclosure is well ventilated Locate the oxygen concentrators low to the ground where it is coolest and where oxygen density is highest Keep in mind that you ll need to look in on your system from time to time It should be easy to access 4 ...

Page 5: ...onfiguration Click here for more information about HVO Power NOTE We recommend against the use of extension cords and power strips Ensure that your power connections are located within 5 feet of your HVO system 5 Master 5 amps Drone naked or dressed 5 amps 3 Outlet Relay Box 20 amps 5 Outlet Relay Box 30 amps ...

Page 6: ...t an Approved Oxygen Regulator Use an oxygen regulator with a flow rate That matches your usage If you have questions about the right Regulator for your setup give us a Call Nut size is about 1 25 CGA standard 6 ...

Page 7: ...ith Phillips Head Bit to open the crate Socket Set To remove tank from crate base Two 18 adjustable wrenches To gently connect O2 regulator Dish Soap Solution in a Spray Bottle to bubble test for leaks Zip ties to secure power cords and oxygen lines 7 ...

Page 8: ...level surface Unscrew and remove the lid 1 Unscrew and Remove the front panel 2 Using the proper socket size remove the Bolts from the base of the tank Lift the tank off the base and place on the floor Carefully remove the remaining contents 2 1 8 ...

Page 9: ...valve is in the open position the tank MUST NOT be under pressure remove the hex nut cap and install a bushing in the opening The bushing will be used to connect the braided steel line that runs between The master and Drone tanks 9 NOTE Don t plug the system in until the last step No power is needed yet side view Remove hex nut cap ...

Page 10: ... the braided steel line to the two tanks The steel line is shipped with teflon tape on both ends Do not install without teflon tape or the fitting will leak oxygen Ensure that the opening in each tank has a bushing into which you can thread the ends of the steel line Use two adjustable wrenches to tighten each End of the steel line One to stabilize the bushing and the other to tighten the fitting ...

Page 11: ...gen manifold lines shown in green below from set A hand tighten the brass fittings to a push to connect manifold black plastic block that terminates with a 3 8 line Repeat for set B Position two relay boxes A B near the tanks Plug in the oxycon power cords from set A to relay box A Repeat for set B Step 8 Prepare the Oxygen Concentrators 11 Set A Set B Box A Box B ...

Page 12: ...nnect in the back of the master Repeat for set B into the naked drone s push to connect Use an adjustable wrench to Connect the oxygen regulator to the master ball valve assembly Do not over tighten Double check all connected components Close the ball valve in preparation to pressurize your system 12 ...

Page 13: ...ug into household power Attach power to the relay boxes plug each relay box into a separate dedicated circuit Plug the naked drone power into one of the relay boxes Attach a coaxial cable from master to relay A Connect a second coaxial cable from relay A to relay B Connect a third coaxial cable from relay box B to the drone 13 ...

Page 14: ...Your system should look like this 14 Drone Set A Box A 14 Set B Box B Rear View Power Oxygen Coaxial Key 20 AMP 30 AMP Master O2 Assumes 6 oxycons Your power requirements may differ ...

Page 15: ...should come to life You should see the lights on the oxycons turn on and you ll hear the compressors running in both the master and the drone Watch the analog pressure gauge In a few minutes you should see pressure building up Once there is visible pressure begin testing all fittings for leaks using a sprayer with soapy water If you find a leak turn the system off Drain the tanks fix the leak and ...

Page 16: ...30 psi or 100 psi depending on the model the system will automatically turn on and begin recharging your tanks with oxygen You can leave your system on 24x7 Switch off and close the ball valve when the system will be unused for a period of days Please Remember to mail in your warranty card If you have any questions or issues please call 402 476 0555 or email support highvolumeoxygen com Enjoy your...