ID7000 Series Smart Code Reader User Manual
36
Timeout Value
Timeout value refers to the maximum running time of algorithm, and its unit is ms. The code
reader will stop parsing the images and return results if the time is exceeded the waiting time
configured.
Algorithm Running Mode
It is used to be set the algorithm operating mode. It includes
High Speed (0)
,
High Performance
(1)
, and
Balance (2)
.
High Speed
focuses on recognition speed, and the algorithm can recognize
the code rapidly, while
High Performance
refers to the algorithm can recognize the code that has
distortion, spot or white gap, but its recognition speed is slow.
Balance
refers to the algorithm
makes a balance between speed and performance.
2D Code Max. Size
It refers to the max. recognizable code width. The 2D code will not be recognized if its width
exceeds the configured value.
Mirror Mode
It is useful when the recognized image is a mirror one, mirroring in X coordinate. 3 modes are
available:
Non Mirror (0)
,
Mirror (1)
, and
Adaptive
(2)
.
Downsampling Level
It refers to the pixel sample size that the code reader takes. Increasing this parameter will improve
the code reading efficiency at the expense of code recognition rate.
Note
Increasing this parameter value will improve the code reading efficiency at the cost of code
recognition rate.
Code Color
It defines the readable code color.
Black Code On White Wall (0)
means that the client software
can recognize the black code with white background.
White Code On Black Wall
(1)
means that
the client software can recognize the white code with black background.
Adaptive (2)
means that
the client software can recognize both the black code with white background, and the white code
with black background.
Note
●
For QR code, the code color is determined by the color of the concentric square on it.
indicates that the code color is white, and
indicates that the code color is black.