4.14 Column Stand Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 24
TRANS
l
LIFT
TM
Resident Stand Assist Service Manual (man173)
4
4.14 Column Stand Assembly
Tools required:
3/8" open end wrench
Removal
1. Disconnect the control box wiring harness.
2. Remove all attachments to the column stand assembly.
3. Remove the two hex head bolts (A) and two flatwashers (B) from the
column stand assembly (C) (see figure 4-15 on page 4-24).
Figure 4-15. Column Stand Assembly Removal
4. Remove the column stand assembly (C) from the base (D).
Replacement
1. Place the two pendant storage labels (E) and the data plate label (F) onto
the column stand assembly (C).
2. Perform the removal procedure in reverse order.
m173_058